Page 18 of 34 

SUPER SESI, by FIT 

User manual, Revision 2.0  

 

Changing the electrospray silica capillary 

CAUTION

  During  this  operation,  the  electrospray  voltage  must  be  set  to  zero,  the 

pressure of the electrospray vial holder must be set to zero, and the temperature of the 
ionization  chamber  must  be  set  below  40ºC.  Once  the  ionization  chamber  set 
temperature is changed, wait for at least 30 min until it cools down. 

If the spray is on, set the electrospray voltage to zero. 

See section ‘Stopping the electrospray´ 

for more details. 

If the MS is acquiring, set the MS to stand-by mode. Refer to the user manual of your mass 
spectrometer for this. 

If  the  vial  holder  is  pressurized,  depressurize  it.  For  this,  make  sure  that  the  syringe  is  not 
connected to the syringe port and open the syringe valve 2 (up position). 

If the ionization chamber temperature is high, set it to a safe value and wait for it to cool down. 

IMPORTANT  NOTE

  It  is  very  important  to  make  sure  that  the  silica  capillary,  the 

electrospray  probe  and  the  electrospray  vial  holder  are  free  from  contamination. 
Always use gloves when manipulating them, and make sure that all surfaces are very 
clean and free from any contamination that could affect your measurements. If these 
parts get contaminated by accidental contact with the skin or with other contaminated 
surface, the background of your measurements will be affected until you clean them.  

Once  the  system  is  cooled,  the  voltage  is  switched  off,  and  the  vial  holder  is  depressurized, 
unlock the electrospray probe, gently pull it out from the ionization chamber, and put it on the 
maintenance position. 

 

Figure 14. Unlocking the electrospray probe . 

*The figure may differ from the latest 

design.  

TIPS & TRICKS

 In order to reduce the pressure needed to pressurize the vial holder and create 

the Electrospray, we recommend to use a sharp silica capillary (20um id) as short as 
possible*, and an extension of silica capillary (50um id) unto the vial holder. For this, we   
provide a Union fitting that is attached to the Electrospray Probe Handgrip, so the two 
capillaries described previously can be connected at this point. 

*

 This length must be long enough to reach the Fitting union (Figure 15). 

Содержание SUPER SESI

Страница 1: ...SUPER SESI Airborne Molecules Metabolites Analysis User Manual Revision 2 1 February 2020...

Страница 2: ......

Страница 3: ...this document is for reference purposes only The system configurations and specifications in this document supersede all previous information received by the purchaser This document is not a sales con...

Страница 4: ...Page 2 of 34 SUPER SESI by FIT User manual Revision 2 0...

Страница 5: ...ion cover 11 Components Nano amperemeter 11 Components nano amperemeter charging unit 12 Symbols used in the SUPER SESI 12 Technical specifications 13 Installation 14 Installation requirements 14 Unbo...

Страница 6: ...e Series Orbitrap Exactive series LTQ Series LCQ Fleet Series Orbitrap LTQ Orbitrap Elite We are working to increase the number of compatible mass spectrometers If you have instrument compatibility qu...

Страница 7: ...e make sure that you understand the special notices symbols and caution labels in this guide Most of the special notices and cautions appear in boxes those pertaining to safety also have corresponding...

Страница 8: ...nded electrospray liquid is high purity water with 0 1 formic acid but other solvents can also be used including flammable solvents such as methanol or ethanol Use care when handling these solvents Al...

Страница 9: ...ntify specific responses to controlled stimuli and to differentiate confounding variables from real discoveries SUPER SESI incorporates all the components and controls in one single unit It is ready t...

Страница 10: ...Page 8 of 34 SUPER SESI by FIT User manual Revision 2 0 Complete system SUPER SESI Figure 1 SUPER SESI front view...

Страница 11: ...User manual Revision 2 0 Components Main Body The main body incorporates the ionization chamber the sample line all the controls and the coupling with the mass spectrometer Figure 2 Main body front vi...

Страница 12: ...corporates a positioning system that allows repeatable and optimum conditions It closes the ionization chamber and guides the sample gas smoothly to the electrospray tip to minimize convective evapora...

Страница 13: ...gure 6 Sample inlet protection cover Components Nano amperemeter The nano amperemeter provides a measurement of the electrospray current This parameter and the current measured by the mass spectromete...

Страница 14: ...ls used in the SUPER SESI Label Name Label design Location Purpose Identification label Besides power connector Identification Contact us Rear panel Contact info for technical assistance General warni...

Страница 15: ...ample line temperature controller Identify controller Valve V1 identifier Near valve V1 Identify valve V1 Valve V2 identifier Near valve V2 Identify valve V2 Syringe inlet identifier 1 Near syringe in...

Страница 16: ...ESI The SUPER SESI is packed with the nano amperemeter the vial holder the electrospray probe and the Mass Flow Controller in their working position In order to protect the SUPER SESI from contaminati...

Страница 17: ...nstall the SUPER SESI on the mass spectrometer in the same way as the original Thermo ion source For this Carefully align the two guide pin holes on the rear of the SUPER SESI with the ion source hous...

Страница 18: ...uid and the vial holder are free from contamination Always use gloves when manipulating them and make sure that all surfaces are very clean and free from any contamination that could affect your measu...

Страница 19: ...wiping it with a Kimwipe soaked with new charging liquid Figure 13 Changing the electrospray liquid Clean the vial or use a new clean vial with new liquid The maximum level of liquid shall be 3 4 par...

Страница 20: ...pray probe and the electrospray vial holder are free from contamination Always use gloves when manipulating them and make sure that all surfaces are very clean and free from any contamination that cou...

Страница 21: ...hard materials Refer to the instructions provided with the silica capillaries for handling The new 20 m capillary has to be first inserted and locked in the electrospray probe For this first make sur...

Страница 22: ...lary positioner align the reference sign of the capillary positioner and the electrospray probe handgrip and lock it in this position Then unlock the capillary lock and the Upchurch fitting 1 they sho...

Страница 23: ...ch must be in the ON position Figure 19 SUPER SESI main switch IEC C14 connector and temperature controllers The SUPER SESI incorporates two independent Proportional Integral Derivative PID temperatur...

Страница 24: ...temperature to be reduced before touching any surface after reducing the set value to safe temperatures CAUTION The temperature control settings are locked to make sure that the SUPER SESI is operated...

Страница 25: ...a is measured 1000 times every second but refreshed more slowly on the display Press once the button above or below of any of the four flow parameters shown to highlight its value in the center of the...

Страница 26: ...flows 2 Make sure that the setpoint at the MFC right up value is higher than the flow value read at the MFC center value to ensure that the clean gas is not restricted by the MFC 3 To fine tune the cl...

Страница 27: ...sed to align the internal absolute pressure sensor with the barometric pressure reading When auto tare is ON your flow controller automatically tares its flow rate whenever it has a zero setpoint for...

Страница 28: ...ulator Note that the high pressure circuit has some inertia You will see that the pressure display rises slowly when the regulator is actuated until it reaches a plateau Once the pressure is raised th...

Страница 29: ...an the sample line between measurements To adjust the clean gas refer to section Adjusting the clean gas flow and the sample gas flow CAUTION Hot Surface At operating temperatures the sample line can...

Страница 30: ...to remove this contamination it is advisable to rise the temperature from time to time so as to evaporate these contaminants and to sweep them away If the spray is on set the electrospray to zero See...

Страница 31: ...as Kimwipes Cotton tipped applicators MeOH H2O HPLC grade and MeOH HPLC grade Tweezers and a tray Allen wrenches MS Syringe pump Cleaning the vapor path In the event that the SUPER SESI gets contamina...

Страница 32: ...e soaked with the cleaning liquid CAUTION When cleaning the vapor path with cleaning liquid the SUPER SESI must not be connected to the mass spectrometer Solvents entering accidentally in the mass spe...

Страница 33: ...lated and cannot be replaced When the batteries run out of charge you will need to recharge them Disconnect the Nano amperemeter from the SUPER SESI by pulling it gently and connect the nano amperemet...

Страница 34: ...system cools down and restart the power If the problem persists contact FIT for technical assistance Replacing the fuses In order to change a fuse first make sure that the power cord is not connected...

Страница 35: ...Fossil Ion Technology S L www fossiliontech com Calle la Gitanilla No 17 Nave 22 ZIP 29004 Malaga Spain email info fossiliontech com...

Страница 36: ...SUPER SESI for Thermo User manual 2019...

Отзывы: