DIGITAL AUTOMATICBLOOD PRESSURE MONITOR
USER MANUAL
MODEL : MD41W0/MD41W1
Страница 1: ...DIGITAL AUTOMATIC BLOOD PRESSURE MONITOR USER MANUAL MODEL MD41W0 MD41W1...
Страница 2: ...hygmomanometers Part 3 Supplementary requirements for electro mechanical blood pressure measuring systems EN 60601 standard relating to medical electrical equipment Part 1 2 General requirements for b...
Страница 3: ...o their condition the oscillometric measurement method can produce incorrect readings This represents a risk for your health since values may be interpreted incorrectly Always consult your physician t...
Страница 4: ...for different classifications of blood pressure If you notice abnormal or suspicious variations in blood pressure measurements consult your physician immediately Women who underwent a breast or axilla...
Страница 5: ...ion to keep a leaking battery away from fire as there is a risk of ignition or explosion In case battery fluid leaks and come into contact with your eyes or skin do not rub and immediately rinse with...
Страница 6: ...kin contact e g put on protective gloves and clean the battery compartment with dry cloth Do not disassemble the device it may cause damage to yourself and to the product If you cannot fix the problem...
Страница 7: ...ver attempt to repair open and or disassemble the unit or adjust it yourself This may cause damage to the unit and impair functions If you cannot fix the problem using the TROUBLESHOOTING instructions...
Страница 8: ...clock mode if idle for 2 min Recalling average data Press button Select between different users User 1 User 2 User 3 User 4 Press or button Air socket Start Stop Button Increase Button Decrease Button...
Страница 9: ...Previous Measurement Memory WIFI Transmission Icon Communicating Icon Low Battery Indicator Irregular Heartbeat Indicator Diastolic Pressure Systolic Pressure Pulse Rate Severe Hypertension Moderate H...
Страница 10: ...9 CLOCK MODE STANDBY MODE Clock Mode Standby Mode...
Страница 11: ...d match the making on the battery compartment 4 Place back the battery door NOTE Insert the batteries as shown in the battery compartment If not the device will not work or even be damaged When batter...
Страница 12: ...t the adapter plug from the AC outlet then the adapter cord from the device NOTE Optional AC adaptor should comply with the requirement of IEC 60601 1 standard Use only the exclusive AC adapter specif...
Страница 13: ...icon is displayed and will blink 13 Press button to select WIFI on or off 14 Set the WIFI to On mode press and hold and buttons for 3 seconds to enter AP mode 15 AP is shown and WIFI icon will blink 1...
Страница 14: ...utton to confirm 3 When On icon is selected Hour will blink Press button to adjust to desired hour Press M button to confirm 4 Minute will blink Press button to adjust to desired minute Press M button...
Страница 15: ...monitor will not be able to take proper measurement Blood circulation in the arm should not be restricted by tight clothing or other objects 3 Pull the cuff so that the top and bottom edges are fitted...
Страница 16: ...you are excited by emotional stress or muscular movement the measurement result may be inaccurate You should not be physically tired or exhausted while taking measurement Perform measurements in a qu...
Страница 17: ...cuff starts to inflate It is normal for the cuff to feel very tight A pressure bar indicator is displayed during measurement See PRESSURE BAR INDICATOR for more details 5 Once the pulse is detected th...
Страница 18: ...17 PRESSURE BAR INDICATOR The indicator monitors the progress of pressure during measurement Pressurizing Measurement in progress...
Страница 19: ...NDICATOR Each of the six segments of the bar indicator corresponds to the WHO blood pressure classification WHO Classification Indicator Severe Hypertension Moderate Hypertension Mild Hypertension Hig...
Страница 20: ...utton in standby mode to enter the memory mode the average blood pressure for the latest 3 measurements is displayed 2 Press button to view average data for lastest 7 days AM record 5 00 9 00 am 3 Pre...
Страница 21: ...n the device is showing average or previous measurement data of selected user 2 Press and hold and for 3 seconds until CL and 00 are shown 3 All measurement data for the selected user will be deleted...
Страница 22: ...ressure monitor Giving voice commands When using the skill you need to specify the name which the blood pressure monitor use and also specify the user ID Use the blood pressure monitor name s you set...
Страница 23: ...alk during measurements NOTE We recommend you to contact your physician if you see this indicator frequently The irregular heartbeat function does not replace a cardiac examination but may help to det...
Страница 24: ...l Do not smoke Exercise regularly Reduce salt and fat intake Have regular physical checkups Maintain proper weight Why Measure Blood Pressure At Home Blood pressure measured at a clinic or doctor s of...
Страница 25: ...mpartment Loose in plug or contact with outlet IF AC adaptor is used Check the wiring to make sure plug outlet are properly secured ERROR code 1 E1 appears The cuff position is not fastened properly F...
Страница 26: ...mining your blood pressure There is a measuring Error Sit comfortably and still Fasten the cuff again carefully ERROR code 5 E5 appears Cuff over inflated The measurement range is over 300 mmHg It is...
Страница 27: ...r Input 100 240V 50 60 Hz Output 6V 600mA service life 5000 times Operating Temperature Humidity 5 C to 40 C 15 93 RH maximum Storage Temperature Humidity 25 C to 70 C up to 93 RH maximum Operation st...
Страница 28: ...rer Type BF Device cuff and tubing are designed to provide special protection against electrical shocks SYS Systolic Blood Pressure in mmHg DIA Diastolic Blood Pressure in mmHg PUL Pulse EC Directive...
Страница 29: ...energy only for its internal function Therefore its RF emissions are very low and are not likely to cause any interference in nearby electronic equipment Conducted and radiated RF EMISSIONS Compliance...
Страница 30: ...s should be at levels characteristic of a typical location in a typical commercial or hospital environment 50 Hz 30 A m 60 Hz 30 A m Electrostatic discharge IEC 61000 4 2 Rated power frequency magneti...
Страница 31: ...n environment Portable and mobile RF communications equipment should be used no closer to any part of the Sphygmomanometer MD41W0 MD41W1 including cables than the recommended separation distance calcu...
Страница 32: ...nds between 0 15 MHz and 80 MHz are 6 765 MHz to 6 795 MHz 13 553 MHz to 13 567 MHz 26 957 MHz to 27 283 MHz and 40 66 MHz to 40 70 MHz The amateur radio bands between 0 15 MHz and 80 MHz are 1 8 MHz...
Страница 33: ...ed below according to the maximum output power of the communications equipment Rated maximum output power of transmitter W Separation distance according to frequency of transmitter m 10 V m 0 01 0 1 1...
Страница 34: ...measures Reorient or relocate the receiving antenna Increase the separation between the equipment and receiver Connect the equipment into an outlet on a circuit different from that to which the receiv...
Страница 35: ...ste correctly at the end of the usage Please follow Local Ordinances or Regulations for disposal MADE IN CHINA P N 83 M41W0 SEN00A R 1017 0123 Shanghai International Trading Corp GmbH Hamburg Eiffestr...