Before you begin your actual pipetting work, fill and empty the tip 2-3 times with the solution 

that you will be pipetting. Hold the pipette in an upright position while aspirating liquid. The 

grippy should rest on your index finger. Make sure that the tips, pipette and solution are at the 

same temperature.

FORWARD TECHNIQUE

Fill a clean reagent reservoir with the liquid to be dispensed.

1.  Depress the push button to the first stop.

2.  Dip the tip under the surface of the liquid in the reservoir to a depth of about 1 cm and 

  slowly release the push button. Withdraw the tip from the liquid touching it against the 

  edge of the reservoir to remove excess liquid.

3.  Deliver the liquid by gently depressing the push button to the first stop. After a delay of 

  about one second, continue to depress the push button all the way to the second stop. 

  This action will empty the tip.

4.  Release the push button to the ready position. If necessary, change the tip and 

  continue pipetting.

REVERSE TECHNIQUE

The reverse technique is suitable for dispensing liquids that have a high viscosity or a tendency 

to foam easily. The technique is also recommended for dispensing very small volumes. Fill a 

clean reagent reservoir with the liquid to be dispensed.

1.  Depress the push button all the way to the second stop.

2.  Dip the tip under the surface of the liquid in the reservoir to a depth of about 1 cm, and 

  slowly release the push button.

  This action will fill the tip. Withdraw the tip from the liquid touching it against the edge of 

  the reservoir to remove excess liquid.

3.  Deliver the preset volume by gently depressing the push button to the first stop. 

  Hold the push button at the first stop. Some liquid will remain in the tip and this should not

  be included in the delivery.

4.  The remaining liquid should either be discarded with the tip or pipetted back into the

  container.

REPETITIVE TECHNIQUE

The repetitive technique offers a rapid and simple procedure for repeated delivery of the same 

volume. Fill a clean reagent reservoir with the liquid to be dispensed.

1.  Depress the push button all the way to the second stop.

2.  Dip the tip under the surface of the liquid in the reservoir to a depth of about 1 cm, and 

  slowly release the push button. This action will fill the tip. Withdraw the tip from the liquid 

  touching against the edge of the reservoir to remove excess liquid.

3.  Deliver the preset volume by gently depressing the push button to the first stop. 

  Hold the push button at the first stop. Some liquid will remain in the tip and this should not

  be included in the delivery.

4.  Continue pipetting by repeating steps 2 and 3.

Содержание Finnpipette II

Страница 1: ...1 INSTRUCTIONS FOR USE Digital Variable Fixed Volume Fisherbrand Finnpipette II 2000 PARK LANE Pittsburgh PA 15275 800 766 7000...

Страница 2: ...2 A C B 6 A C B 1 2 3 4 4 A C B 7 A C B 5 1 2 11 9 10 3 1 2 3 4 1 2 3 4 1 2 3 4 5 12...

Страница 3: ...3 14 A B C 13 8 CONTENTS PRODUCT DESCRIPTION 4 PIPETTE OPERATION 5 PIPETTING TECHNIQUES 5 CALIBRATION 7 MAINTENANCE 9 TROUBLE SHOOTING 11 PACKAGE 11 SPARE PARTS 12 13...

Страница 4: ...0 200 Ext 21377818 5 l 50 l 250 Universal 300 200 Ext 21377819 10 l 100 l 250 Universal 300 200 Ext 21377820 20 l 200 l 250 Universal 300 200 Ext 21377821 100 l 1000 l 1000 1000 Ext 21377822 200 l 100...

Страница 5: ...gearing mechanism To release the tip point the pipette at suitable waste receptacle and press the tip ejector with your thumb SAFETY LABEL You can mark the pipette application your initials the calib...

Страница 6: ...with the liquid to be dispensed 1 Depress the push button all the way to the second stop 2 Dip the tip under the surface of the liquid in the reservoir to a depth of about 1 cm and slowly release the...

Страница 7: ...graduation value of the balance should be chosen according to the selected test volume of the pipette Volume range readable graduation under 10 l 0 00 1 mg 10 100 l 0 01 mg above 100 l 0 1 mg Test liq...

Страница 8: ...trap or lid on the vessel is used the correction of evaporation is usually unnecessary Range Volume Accuracy Precision l l s d l cv 0 2 2 l 2 0 080 4 0 040 2 0 0 2 0 080 40 0 020 20 0 0 5 10 l 10 0 12...

Страница 9: ...ORT TERM CHECKING The pipette should be checked at the beginning of each day for dust and dirt on the outside surfaces of the pipette Particular attention should be paid to the tip cone No other solve...

Страница 10: ...to the instructions DISASSEMBLING 50 1000 l 41 1000 l PIPETTES 1 Press the tip ejector 2 Insert the pliers under the ejector bar to release the tip ejector 3 Remove the tip cone using the maintenance...

Страница 11: ...between tip and tip cone Foreign particles between the piston the O ring and the cylinder Insufficient amount of grease on cylinder and O ring O ring damaged Incorrect operation Tip incorrectly attach...

Страница 12: ...7 15 15 13 13 14 15 16 13 16 15 14 14 5 17 16 17 10 10 1 2 1 5ml 2 5ml Fix 2 10ml 6 10ml Fix 0 2 2 l 1 2 l Fix 5 50 l 11 40 l Fix 2 20 l 10 100 l 14 15 13 14 16 17 13 14 15 16 20 200 l 41 200 l Fix 10...

Страница 13: ...13 13 13 14 14 12 10 10 20 10 12 12 11 17 18 21 20 19 11 22 23 21 19 18 18 19 20 10 10 11 11 11 12 12 11 12 12 a b 10 10 11 11 11 12 12 19 18 b 20 15 16 15...

Страница 14: ...1131800 20 10593360 21 1030170 Shelf hanger 2206740 10 10593410 11 10593630 12 10593100 13 10589450 14 1130560 15 1130550 16 1054260 17 1030020 20 200 l 41 200 l Fixed 5 10592520 5 10592540 Fix 50 l...

Страница 15: ...0034 1 0035 1 0035 1 0036 23 5 1 0034 1 0035 1 0035 1 0036 1 0036 1 0036 1 0037 24 0 1 0035 1 0036 1 0036 1 0037 1 0037 1 0038 1 0038 24 5 1 0037 1 0037 1 0038 1 0038 1 0039 1 0039 1 0039 25 0 1 0038...

Страница 16: ...16 1507910 07 For customer service call 1 800 766 7000 To fax an order use 1 800 926 1166 To order online www fishersci com 2002 Fisher Scientific Litho in U S A BN...

Отзывы: