All controls and indicators are located on the front panel for ease of
operation.
Power:
Controls line power to unit. Set to 1 position to turn on power, 0 posi-
tion to turn off power.
Temperature Control:
Serves as the primary temperature control for the bath. Turns the heater
on or off to maintain the bath at the set temperature.
Limit Control:
Sets the backup temperature to the primary setpoint; should be set at a
slightly higher temperature than the Temperature Control setpoint to
give safety control in the event that the primary control fails.
Temperature Control Indicator:
The green indicator illuminates to show when heat is being supplied to
the chamber; cycles on and off as heat is requester by the controller.
Limit Control Indicator:
The amber indicator illuminates to show the the Limit Control is regu-
lating the temperature rather than the primary Temperature Control.
Note
If the bath has been used previously, the tank should be thoroughly
cleaned before refilling.
s
Note
The water bath, full of water (within 1-1.5" of top of tank), with lid,
is designed to reach 37°C in less than 60 minutes or 90°C in less than 180
minutes.
s
4-4
Isotemp Economy Water Bath
Fisher Scientific
Section 4
Unpacking and Installation
Controls and
Indicators