7
Installing and Leveling Casters
Isotemp Plus refrigerators and freezers come with four
casters, which thread into the base of the unit, one in
each corner. Use the wrench provided to thread the cast-
ers completely into the base of the unit. Back the casters
in or out until the unit is level and resting on all four cast-
ers. See Figure 1.
Fluorescent Lamps
Prior to applying power to the unit, verify that all of the
internal lamps are properly installed and fully engaged in
the lamp holders. Proper lamp engagement is achieved
by rotating the lamp 90° from its insertion position until it
“snaps” or “clicks” into place.
Since the lamps in this unit are protected by shields, it is
necessary to physically verify that the lamp has been
rotated and locked into place.
Door Removal
(Sliding Door Models Only)
Each door has its own closing spring located at the top of
the door track. Each spring is set for proper tension. If
doors do not close properly:
1.
Check for cabinet level (see above).
2.
Remove the door(s) by opening slightly then lift-
ing it. While maintaining upward pressure on
the door, pull the bottom outward until it clears
the bottom track.
3.
Check that spring in top track moves freely and
that bottom track is free of debris. Check that
rollers on bottom of door rotate freely.
U
NPACKING AND
I
NSTALLATION
Caution
If the unit was tilted in excess of 30
degrees, level it and do not apply
electrical power for a minimum of
12 hours.
Refrigerator
Base
Thread casters directly
into base at four corners.
Tighten using
wrench provided.
Caution
Improperly installed lamps may cause
damage to the lighting circuit.