Model 228 Centrifuge User Trouble Shooting Guide 

Symptom 

Possible Cause 

Possible Solution 

Unit will not run 

No power to unit 

Check that unit is plugged in 
Check that outlet has power 

Rotor loaded improperly 

Check that rotor loading is balanced. 

Rotor not tightened properly. 

Check that rotor fits securely on motor 
shaft and that thumb screw is tight. 

Unit vibrates 

Motor mounting nuts are loose 

View bottom of unit and check that the 
four nuts that hold the motor to the base 
casting are tight. 

See “Unit  vibrates” section 

See “Unit  vibrates” section 

Unit is noisy 

Lid not latched securely 

Check that lid is latched.  If lid is latched 
but not tight, adjust lid latch pawl. 

 
 

Repair Trouble Shooting Guide 

Symptom 

Possible Cause 

Check 

Parts  

Wiring 

Check wiring connections. 

 

Timer 

Check for power in and out of
timer. 

Timer; 106868 

Lid Interlock 

Check that lid latch is 
actuating interlock switch.  
Check that switch is not 
jammed. Check switch 
continuity.  

Leaf Switch;  67516 
 

Unit will not run 

Motor 

Check for power to motor. 
Check brushes.  Check that 
motor shaft turns freely. 

Motor; 50782 (115V), 
50990 (230V),  
 

See User Trouble 
Shooting Guide 

 

 

Damaged Motor 

Check that motor shaft turns 
freely and does not “wobble”. 
Check for shaft “side play”.   

Motor; 50782 (115V), 
50990 (230V),  
 

Unit vibrates or is noisy 

Damaged Rotor 

Check that rotor fits securely 
to shaft and does not 
“wobble”.. 

Rotor;  50779 
Rotor with Shields;  
04-978-233 

 

 
 
 
 
 
 
 

Fisher Scientific Co. 
2000 Park Lane    
Pittsburgh, PA  15275 

Technical Support 1-800-926-0505 
Repair Service: 1-800-395-5442 

Sales: 1-800-766-7000 

Part Number 70906  Rev. F 
Operation & Service Instructions 
Centrific® Centrifuge 04-978-228 & -229 

Published 03/2004

 