![Fisher Bioblock Scientific Wheaton Biostir Скачать руководство пользователя страница 9](http://html1.mh-extra.com/html/fisher-bioblock-scientific/wheaton-biostir/wheaton-biostir_manual_2296473009.webp)
2004 Wheaton Science Products
9
P/N 50083244
Specifications subject to change without prior notice
5.0
INITIAL INSPECTION
When you receive your new Biostir
, inspect it for any obvious damage that may
have occurred during shipment. If any damage is found, notify the carrier at
once. Warranty information is shown in the front of this manual. Check to
confirm that there are no broken switches, displays or missing parts, and that the
unit is not dented or scratched.
5.1
Installation
Install the unit where there will be adequate room for the unit to operate. Provide
enough clearance around the unit to maintain safety. Install the unit on a level
surface to make sure vessels don’t slide off the top.
5.2
Input Power Requirements
This equipment is designed to operate from a 200-240V single-phase AC power
source at 47 to 63 Hz. The line voltage / fuse label located on the upper rear of
the unit shows the nominal input voltage set for the unit at the factory.
5.3
Power Cord Set
This unit has been shipped from the factory with a power line cord that has a plug
appropriate for your area. If the wrong power cord has been shipped for your
particular application, contact your nearest Wheaton Science Products (WSP)
dealer for the proper cord. The Biostir
has been equipped with a 3-wire
grounding type power cord. The unit is only grounded when it is plugged into an
appropriate receptacle. Do not operate the unit without adequate grounding
protection.