FISCHER MD-1a Скачать руководство пользователя страница 8

Electrical Device Inspection

 (Continued) 

Period: Annually (12 months) 
 

Device Output: 
 

PRELIMINARY SETUP:

 

• 

Plug line cord into a grounded receptacle and press the bottom edge of the POWER 
switch to turn on the MD-1a (red portion of power switch will be visible). Note that the 
red  “ACTIVE”  LED  will  NOT  be lit  unless  a patient  circuit is  completed  between  the 
Output Jacks and current is flowing.  

• 

Attach  the  dual  patient  cordset  to  the  Output  Jacks  and  make  all  measurements 
between the red and black banana plugs at the end of the cordset using a series load 
resistance  of  1K  ohm  rated  at  3  watts  or  more.  (A  2  watt  resistor  may  be  used  for 
short duration measurements). 

 
 

OUTPUT MEASUREMENTS

 

Measure Minimum output using a DC milliameter in series with a 1K load resistance 
and  Intensity  control  knob  set  to  its  minimum  position.  This  measurement  simply 
confirms  that  there  is  no  unintended  output  current  with  the  Intensity  knob  at  its 
minimum (zero) position. 

 

 

 

 

Accept: 0 DCma

 

 

Measure Maximum output using a DC milliameter in series with a 1K load resistance 
and  Intensity  control  knob  set  to  its  maximum  position.  Note  that  the  MD-1a  can 
produce approximately 50 DCma at this this setting, which will require a resistor rated 
at 2.5 Watts or higher.   

 

 

 

 

Accept: 40-58 DCma

 

 

Verify  meter  reading  on  0-10  scale.  With  a  1K  load  resistance,  set  the  METER 
SCALE switch to the 0-10 position and adjust Intensity until the the meter settles on a 
reading  of  10ma  (full  scale).  Measure  the  actual  current  using  a  DC  milliameter  in 
series with the load resistance.  

 

 

 

Accept: 8.0-12 DCma

 

 

Verify  meter  reading  on  0-50  scale.  With  a  1K  load  resistance,  set  the  METER 
SCALE switch to the 0-50 position and adjust Intensity until the the meter settles on a 
reading of 20ma. Measure the actual current using a DC milliameter in series with the 
load resistance.   

 

 

 

 

 

Accept: 16-24 DCma

 

NOTE: The MD-1a incorporates circuitry to slowly ramp the output current to its set value 
when  a  load  resistance  is  detected. Therefore,  you  must  allow  several  seconds  for  the 
output current to stabilize before making these measurements. 

 
 

AC RIPPLE (OPTIONAL):

 

 

Measure the AC ripple present in the output current by connecting an 

isolated ground

 

oscilloscope

 

in  parallel  with  the  1K  load  resistance.  Adjust  Intensity  for  a  metered 

output  of  20  DCma  (using  0-50  scale)  and  observe  the  AC  component  (ripple) 
superimposed on the DC waveform.   

 

 

Accept: 0-.5 Vpp (2.5%)

 

 

Note: Output specifications measured at 115VAC line voltage 

DEVICE INSPECTION 

Page 7 

Содержание MD-1a

Страница 1: ......

Страница 2: ...trols Page 3 Device Specifications Page 4 Device Maintenance Page 4 Device Inspection Page 5 Assembly Parts List Page 8 Schematic Diagram Page 9 MODEL MD 1a Galvanic Unit Technical Reference Manual CO...

Страница 3: ...nding although the plug will fit typical grounded receptacles Because of its light weight and small size the MD 1 is very portable When relocating the unit for use however always be sure it is properl...

Страница 4: ...witch will be exposed when the power is turned ON INTENSITY KNOB Adjust this knob to set the desired current level to be applied to the patient The actual current is indicated on the DC MILLIAMPERES m...

Страница 5: ...Patient Override Control N A Indicator Display Unit Functioning Active lamp Current Level analog DC milliameter Meter ranges 0 10 DCma w 0 2ma resolution 0 50 DCma w 1 0ma resolution Meter Accuracy 20...

Страница 6: ...Minimum intensity setting and turn on at a setting of 1 2 on the knob Verify operation of Polarity Reversal switch applies to units made after 1 1 2005 Switch in REV position will reverse polarity at...

Страница 7: ...t Nominal readings appear below for reference Normal Polarity Normal Ground 5 uA Reverse Polarity Normal Ground 5 uA Normal Polarity Open Ground 10 uA Reverse Polarity Open Ground 10 uA Measure patien...

Страница 8: ...pproximately 50 DCma at this this setting which will require a resistor rated at 2 5 Watts or higher Accept 40 58 DCma Verify meter reading on 0 10 scale With a 1K load resistance set the METER SCALE...

Страница 9: ...Resistor 1 00 38 0004 10K 2W 53C3 Pot 1 00 41 0013 100uf 100V Electrolytic 1 00 41 0015 100uf 250V Electrolytic 2 00 41 0033 33uf 25V Tantalum 1 00 45 0015 005uf 50V Ceramic Disc 1 00 45 0006 001uf A...

Страница 10: ...Page 9...

Отзывы: