FISCHER MD-1a Скачать руководство пользователя страница 3

MD-1a Galvanic Unit 

 

The  Model  MD-1a  Galvanic  Unit  offers  a  versatile  approach  to the  use  of  Galvanic  (DC) 
current  in  certain  medical  applications  employing  anaphoresis,  cataphoresis,  or 
iontophoresis.  It  is  designed  for  use  by  physicians  in  a  clinic  setting,  and  generates  an 
adjustable, filtered DC current from normal AC line voltage. Standard accesssories supplied 
with the unit include: 
 

 

 

(1) Dual Patient Cordset (red and black cords) 

p/n 15-0003 

 

(2) 3" x 6" Wet pad electrodes 

 

 

p/n 12-0019 

 

(1) Iontophoresis Trays (OPTIONAL) 

 

p/n 13-0020 

 

 

 

 

TYPICAL INSTALLATION

 

 
Locate the Galvanic Unit near a treatment area where all controls are easily accessible. Plug 
the  line  cord  into  a  110-115  volt  A.C. 

grounded

  wall  receptacle  (220  volt  operation  is 

available on special order only).  

DO  NOT  USE  "CHEATER"  OR  ADAPTER  PLUGS.  PROPER  GROUNDING  IS 
ESSENTIAL FOR SAFE AND PROPER OPERATION OF THIS DEVICE. 

The MD-

1a is equipped with a “hospital grade” plug to ensure reliable grounding, although the plug 
will fit typical grounded receptacles.

 

Because of its light weight and small size, the MD-1 is 

very  portable.  When  relocating  the  unit  for  use,  however,  always  be  sure  it  is  properly 
grounded. 
 
Insert the dual-prong plug at one end of the patient cordset into the front panel jacks; the red 
jack is always positive with respect to the black (negative). The plugs at the opposite ends 
of  each  cord  should  be  firmly  plugged  into  the  accessory  electrodes  to  be  used  for 
treatment.

 

  !   

  

 

PRECAUTIONS 

CAUTION:  PRODUCES  HIGH  VOLTAGE.  Always  exercise  care  and  follow  safety 
procedures when operating or servicing this device. 

Disconnect  the  power  cord  when  servicing  the  unit  to  reduce  the  risk  of  shock  hazard. 
Avoid contact with in-circuit capacitors which may retain a high voltage charge. 

Wait at least 3 minutes after disconnecting from power source to allow circuit components 
to fully discharge before performing any service. 

Do not move or carry the unit by any attached cords. 

Do  not  subject the  machine  to  excessive  shock  or  vibration;  do  not  use or  store  in  places 
where termperature is over 105 degrees, or in high humidity.

 

Page 2 

Содержание MD-1a

Страница 1: ......

Страница 2: ...trols Page 3 Device Specifications Page 4 Device Maintenance Page 4 Device Inspection Page 5 Assembly Parts List Page 8 Schematic Diagram Page 9 MODEL MD 1a Galvanic Unit Technical Reference Manual CO...

Страница 3: ...nding although the plug will fit typical grounded receptacles Because of its light weight and small size the MD 1 is very portable When relocating the unit for use however always be sure it is properl...

Страница 4: ...witch will be exposed when the power is turned ON INTENSITY KNOB Adjust this knob to set the desired current level to be applied to the patient The actual current is indicated on the DC MILLIAMPERES m...

Страница 5: ...Patient Override Control N A Indicator Display Unit Functioning Active lamp Current Level analog DC milliameter Meter ranges 0 10 DCma w 0 2ma resolution 0 50 DCma w 1 0ma resolution Meter Accuracy 20...

Страница 6: ...Minimum intensity setting and turn on at a setting of 1 2 on the knob Verify operation of Polarity Reversal switch applies to units made after 1 1 2005 Switch in REV position will reverse polarity at...

Страница 7: ...t Nominal readings appear below for reference Normal Polarity Normal Ground 5 uA Reverse Polarity Normal Ground 5 uA Normal Polarity Open Ground 10 uA Reverse Polarity Open Ground 10 uA Measure patien...

Страница 8: ...pproximately 50 DCma at this this setting which will require a resistor rated at 2 5 Watts or higher Accept 40 58 DCma Verify meter reading on 0 10 scale With a 1K load resistance set the METER SCALE...

Страница 9: ...Resistor 1 00 38 0004 10K 2W 53C3 Pot 1 00 41 0013 100uf 100V Electrolytic 1 00 41 0015 100uf 250V Electrolytic 2 00 41 0033 33uf 25V Tantalum 1 00 45 0015 005uf 50V Ceramic Disc 1 00 45 0006 001uf A...