Страница 2: ...trols Page 3 Device Specifications Page 4 Device Maintenance Page 4 Device Inspection Page 5 Assembly Parts List Page 8 Schematic Diagram Page 9 MODEL MD 1a Galvanic Unit Technical Reference Manual CO...
Страница 3: ...nding although the plug will fit typical grounded receptacles Because of its light weight and small size the MD 1 is very portable When relocating the unit for use however always be sure it is properl...
Страница 4: ...witch will be exposed when the power is turned ON INTENSITY KNOB Adjust this knob to set the desired current level to be applied to the patient The actual current is indicated on the DC MILLIAMPERES m...
Страница 5: ...Patient Override Control N A Indicator Display Unit Functioning Active lamp Current Level analog DC milliameter Meter ranges 0 10 DCma w 0 2ma resolution 0 50 DCma w 1 0ma resolution Meter Accuracy 20...
Страница 6: ...Minimum intensity setting and turn on at a setting of 1 2 on the knob Verify operation of Polarity Reversal switch applies to units made after 1 1 2005 Switch in REV position will reverse polarity at...
Страница 7: ...t Nominal readings appear below for reference Normal Polarity Normal Ground 5 uA Reverse Polarity Normal Ground 5 uA Normal Polarity Open Ground 10 uA Reverse Polarity Open Ground 10 uA Measure patien...
Страница 8: ...pproximately 50 DCma at this this setting which will require a resistor rated at 2 5 Watts or higher Accept 40 58 DCma Verify meter reading on 0 10 scale With a 1K load resistance set the METER SCALE...
Страница 9: ...Resistor 1 00 38 0004 10K 2W 53C3 Pot 1 00 41 0013 100uf 100V Electrolytic 1 00 41 0015 100uf 250V Electrolytic 2 00 41 0033 33uf 25V Tantalum 1 00 45 0015 005uf 50V Ceramic Disc 1 00 45 0006 001uf A...