background image

5

time-constant (averaging filter) in the user selectable ran-
ge of 0.0 to 100.0 secs. When the damping is set to ma-
ximum,  it  takes  more  than  2  minutes  to  reach  the  final
value for a full scale pressure jump.

In many cases fluctuating pressure readings do not cau-
se  a  problem,  except  when  the  plant  /  equipment  is  at
zero  (differential)  pressure  condition  and  readings  fluc-
tuate near about zero value. The set-up parameter 

NP

 is

meant to take care of this. Its value defines the number
range  across  zero  (similar  to  the  zero  offset  correction
number),  within  which  the  measured  value  is  forced  to
zero. If a value of 8 is set for 

NP

 any pressure measure-

ment  in  the  range -0.08 to +0.08  mbar (or -8 to +8 Pa)
displayed as zero. Only when the actual pressure is out-
side this range will the display indicate a non-zero value.
The  actual  and  displayed  pressures  will  agree  starting
from double the value of the 

NP

 setting (in the given ex-

ample: 0.16 mbar or 16 Pa).

5.3.5. Output Signal Setting

The  transmitter  output  signal  depends  primarily  on  the
measured pressure. However, this signal can be adapted
to  meet  users'  application  requirements  precisely.  The
basic pressure range (as marked on the product identifi-
cation label) and the type of output signal (voltage or cur-
rent)  always  remain  unchanged  for  a  particular
instrument unit. 

The  set-up  parameters 

MA

  (measuring  range  starting

point)  and 

ME

  (measuring  range  end  point)  specify  the

pressure  values  between which  the  measurements  are
expected to be. Both values can selected anywhere wi-
thin  the  specified  measuring  range  of  the  instrument
(e.g.,  400 Pa).  This  user-programmed  pressure  range
will  correspond  to  the output  signal  (current  or voltage)
range, which will be as specified on the product identifi-
cation plate: i.e., 0 - 10 V or 4 - 20 mA. If 

MA

 is lower than

ME

,  the  signal is  said to  have a positive  slope: i.e., the

output signal increases as the pressure increases. If 

ME

is lower than 

MA

, the output signal has a negative slope:

i.e., the output signal decreases as the pressure increa-
ses. 

The difference between the values of 

MA

 and 

ME

 must be

at least 25% of the specified measuring range of the in-
strument  (100 Pa for 400 Pa instrument example men-
tioned  above).  The  software  does  not  permit  a  smaller
pressure span to be entered (the instrument will not allow
storing of, nor exit from an invalid span).

Note: If you change 

MA

 and/or 

ME

 the

look-up table (s .5.3.7., 5.3.8.) that exi-
sted up to that instant is deleted! 

5.3.6. Output Signal Limiting (Namur)

The three set-up parameters 

oG1

oG2

 and 

oEr

 specify

the limits of the signal output current or voltage that are
not to be exceeded, irrespective of the actual pressure.

These limit values have higher priority than the 

MA

 and

ME

 pressure span settings.

These settings serve mainly to prevent control systems
from  interpreting  brief  pressure  excursions  outside  the
measuring range as error / fault events. 

oG1

 sets the out-

put  signal  minimum  value,  and  is  useful  only  for  a
4 - 20 mA current signal, because sometimes a value be-
low  3.8 mA  is  defined  as  a  sensor  fault condition. 

oG2

sets the upper signal limit and is valid for either current or
voltage signal (e.g., voltage signal output can be limited
at 10.2 V).

An  instrument  fault  condition  can  be  transmitted  as  an
output  signal  value  set  as  parameter 

oEr

.  However,  it

should be understood that not all instrument fault and er-
ror conditions can be detected and signaled by the self-
diagnostic functions of the instrument!

5.3.7. Signal Conversion and Transfer Functions

In certain cases other process variables are derived from
primary pressure measurement. Examples are flow rate
derived from differential pressure across an orifice plate,
and  liquid  level  derived  from  hydrostatic  pressure  of  li-
quid measured at the bottom of the tank. Such derivati-
ons often involve non-linear transfer functions, while it is
necessary for the output signal to be linearly proportional
to the derived variable (e.g. liquid volume in a tank in m

3

,

or flow rate in cm

3

/sec). 

The set-up parameter 

F

 allows the user to select the ap-

propriate signal conversion function from those available:

•

F

 = 0: Linear characteristic (default)

•

F

 = 1: Square root extraction

•

F

 = 2: Horizontal cylindrical tank

•

F

 = 3.....30: Look-up table with 3 to 30 pairs of values

The tables generated by functions 

F

 = 0, 

F

 = 1 and 

F

 = 2

are  not  visible.  For  these  functions,  internal  values  are
used for table computation. These values cannot be mo-
dified by the user.

For all conversion functions, when the actual pressure is
equal to the 

MA

 value, the output signal will be at the lo-

west end of its range (0 V, 0 mA or 4 mA). When the ac-
tual pressure is equal to the 

ME

 value, the output signal

will be at the  highest end  of  its  range  (10 V or 20 mA).
The user can enter only the 1...28 intermediate values of
the look-up table function 

F

 = 3....30. The parameters 

MA

and 

ME

 relate to the start and end values of the look-up

table. A change in either of these parameter values cau-
ses the conversion function to be re-set to 

F

 = 0.

Whenever  the  value  of 

F

  is  changed,

the  instrument  internally  generates  a
new  look-up  table.  All  previous  table
values  are  deleted  and  replaced  by
new linear progression values.

Contact: EK Engineering (India)
Web: www.ekeindia.com
Email: [email protected]

Содержание DE44 KR Series

Страница 1: ...tallation operation maintenance and inspection of this product must have the qualifications training and experience necessary to carry out such work on this type of equipment Qualified personnel are people who are able to judge delegated work and possible dangers due to technical education proficiency and experiences and especially by knowledge about the relevant norms When working with explosion ...

Страница 2: ...oduct if and as necessary should be done only by Fischer Mess und Regeltechnik GmbH 1 6 Operational Restrictions The operational reliability of the product is guaranteed only when used for intended purposes The product must be selected and configured for use specifically with defi ned process media The limiting values of operating pa rameters as given in the product specification sheet must never ...

Страница 3: ...hazardous areas e g EN 60079 14 EN 50014 Operating supply voltage 24 V DC AC must not excess 32 V DC AC neither per temporary inter ferences Supply wiring system must be protec ted by Si 315 T For recommended current supply see 11 Specifi cations Parametrisation by PC serial interface adaptor EU03 F300 is permitted only in safe areas outside explosion hazardous areas zone 2 5 Commissioning Power s...

Страница 4: ...s again û to store the selection and EIN will appear again in the digital display To exit the set up mode press ÿ until ESc appears and then press û The current pressure measured value is indicated again and the appropriate symbol of the unit of measurement to the right of the digital display is lit The digital display is limited to a count of 1999 Therefore all the available units of measurement ...

Страница 5: ... oG2 and oEr specify the limits of the signal output current or voltage that are not to be exceeded irrespective of the actual pressure These limit values have higher priority than the MA and ME pressure span settings These settings serve mainly to prevent control systems from interpreting brief pressure excursions outside the measuring range as error fault events oG1 sets the out put signal minim...

Страница 6: ...tput turns on when the measured va lue exceeds R1A R1E and turns off again when the measured value falls below R1A R1E If R1A is larger than R1E the switching output turns on when the measured value falls between R1A and R1E i e when R1E measured value R1A window limit function Both parameters can be independently adjusted over the full measuring range If the unit of measurement is changed the swi...

Страница 7: ...easuring range start point The value of the measured variable corresponding with the mi nimum value of the output signal 0 V 0 mA or 4 mA depending on the instrument version ME Measuring range end point The value of the measured variable corresponding with the ma ximum value of the output signal 10 V or 20 mA depending on the instrument version dPF Position of decimal place for free unit MAF Measu...

Страница 8: ...htness Check the integrity of all electrical connections of the instruments Inspection and test schedules depend on operating and site conditions The operating manuals of other equip ment to which the transmitter is connected must be read thoroughly to ensure that all of them work correctly when connected together 7 Transport The product must be protected against shock and vibra tion during transp...

Страница 9: ...9 11 Specifications 11 1 General Two independent differential pressure sensors Contact EK Engineering India Web www ekeindia com Email india eke gmail com ...

Страница 10: ...e for signal output display seperated Display function 0 0 Pressure port 1 only 0 1 Pressure port 2 only 0 2 25 5 Display alternating time 0 Relay switch 1 2 Activation point de activation point response time delay 0 0 100 0 secs logic N O or N C Measurement unit selection mbar Pa free unit start value end value and decimal place for free unit Zero suppression 0 100 counts 1 Output signal start en...

Страница 11: ... mbar 5 8 0 100 mbar 500 mbar 5 9 2 5 2 5 mbar 50 mbar A 6 4 4 mbar 50 mbar A 7 6 6 mbar 100 mbar A 8 10 10 mbar 100 mbar A 9 16 16 mbar 250 mbar B 1 25 25 mbar 250 mbar B 2 40 40 mbar 500 mbar C 5 60 60 mbar 500 mbar B 3 Pressure Connections Threaded hose clamp connectors for 6 4 mm hose 4 0 Threaded hose clamp connectors for 8 6 mm hose 4 1 Signal Output Pressure Port I only No signal output 0 C...

Страница 12: ...switch 2 m 06401994 Cable with M12 connector 4 pin for switch 5 m 06401995 Cable with M12 connector 5 pin for supply signal 2 m 06401996 Cable with M12 connector 5 pin for supply signal 5 m 04005144 Wall mounting adaptor plate EU03 F300 PC programming interface with SW Contact EK Engineering India Web www ekeindia com Email india eke gmail com ...

Отзывы: