Power Meter Operation

Taking a reference and test: 

Two reference cord method

a. 

First turn on light source and press F1 to run source in Auto 

              Mode (default setting). Let it cycle through for about five 

 minutes.

b. 

 Clean the power meter port (male) as well as the female MPO 

 

connector you are plugging in.  Please do the same for the 

 

Light Source port, which is also male. 

c. 

Make sure you only insert female MPO connectors into the 

 

power meter at all times.

d. 

Once the source has stabilized after five minutes, connect your 

 

Type A polarity reference cords (provided) to the power meter 

 

and light source. Use the provided mating sleeve in the kit to 

 

connect the two reference cords together between the meter 

 

and source (male to female).

e. 

Now, while the light source is sequencing through, turn the 

 

power meter on and confirm that the meter is reading power 

 

(dBm) on each channel as the source sequences through. 

f. 

When ready, press F1 on the METER screen (“Zero Ref”) to take 

 

a power reference

g. 

Follow the on screen instructions…

h. 

Once the reference is complete, disconnect the reference cords 

 

from each other in the middle and introduce the cable under 

 

test in between them. 

i. 

Once the ‘cable under test’ cable is introduced, highlight the 

 

polarity type of the cable you will be testing with the arrow 

 

cursor and press F1 when ready to test.

j. 

The Power Meter will now automatically go back to the METER 

 

screen and run a relative loss test; the value of measurement 

 

will be in dB.  

k. 

Once the source has scanned all 12 fibers, the Meter will have 

 

the loss (dB) values displayed and only if a P/F threshold was 

 

set, a PASS or FAIL will be displayed on the screen along with 

 

the user defined threshold.  If one fiber exceeds the set thresh

 

old the entire test fails.  A “P” represents an individual fiber 

 

passed and an “X” is an individual fiber that failed.

Содержание MPO 12

Страница 1: ...MPOTest Set User Manual...

Страница 2: ......

Страница 3: ...fined attenuation thresholds and will display a FAIL if any of the 12 fibers exceeds the threshold If all fibers are within the threshold a PASS will be displayed This unit can also store up to 1 000...

Страница 4: ...can be set in the options menu under pass fail thresholds I Auto Off This feature is activated as the default setting every time the unit turns on To de activate it short press the on off button and i...

Страница 5: ...2 Fiber Instrument Sales Inc 1 800 5000 FIS 347 www fiberinstrumentsales com A B C D E F G H I J K L M N...

Страница 6: ...hreshold You can set a threshold down to the hundredths of a dB 0 01 Use the left right arrows to choose which placement to control dictated by the underscore _ and then use the up down arrows to the...

Страница 7: ...lluminated H Auto Off For the source this feature is de activated as the default setting every time the unit turns on To activate it short press or tap the on off button and AUTO OFF will appear on to...

Страница 8: ...5 Fiber Instrument Sales Inc 1 800 5000 FIS 347 www fiberinstrumentsales com A B C E D G H K I J L M F...

Страница 9: ...12 channels until you press F1 again You will notice a lightly shaded round icon moving through the channels when the source is active When the icon goes dark the source is off To change the mode at...

Страница 10: ...h turn the power meter on and confirm that the meter is reading power dBm on each channel as the source sequences through f When ready press F1 on the METER screen Zero Ref to take a power reference g...

Страница 11: ...e file number is already taken it will say used if not it will say free You can either overwrite a used file or save into a free spot 5 The unit will always default to the next available free file slo...

Страница 12: ...your PC E You can select the provided default folder or create your own then select the folder F Then the software will tell you to select a COM port to connect the unit with the PC software G Select...

Страница 13: ...enu click Reports Then choose either Generate to convert highlighted file only or Generate All to generate all exported files To generate all files into a report click Generate All This will convert a...

Страница 14: ...Notes...

Страница 15: ...Notes...

Страница 16: ...161 Clear Road Oriskany NY 13424 Ph 315 736 2206 Toll Free 1 800 5000 FIS 347 Fax 315 736 2285 www fiberinstrumentsales com email info fiberinstrumentsales com...

Отзывы: