![First Light C-RED 3 Скачать руководство пользователя страница 32](http://html1.mh-extra.com/html/first-light/c-red-3/c-red-3_user-manual_2293789032.webp)
32
Figure 20:
Windows device manager illustration
The COM port settings are the same as Camera Link
®
serial port interface:
115200 Bauds, 8 bits, No parity, 1 Stop bit, no flow control
A USB SDK for windows 10 and Linux is provided in both binary and source forms to perform frame
grabbing.
Proper operation of the camera has been tested when using common one-meter long USB 3.0 cables,
and with up to five-meters industrial grade USB 3.0 cables.
Internal synchronization
In this mode, the camera generates frames according to the configured acquisition framerate. Each
image is integrated during the configured integration time.
This is the default configuration of the camera.