13.3 Error messages
Fault
Possible cause
Troubleshooting
Responsibility
"Er00" displayed
Door open during
process
Close door (“Press”)
Poss. readjust door
locking
User
Service
department
FINO
“Er01" displayed
Excessive tempera-
ture
Replace the
themocouple.
Service
department
FINO
"Er02" displayed
Broken sensor, mea-
suring circuit
Retighten the ther-
mocouple connec-
tions.
"Er03" displayed
Broken sensor
Replace the ther-
mocouple.
“Er04" displayed
Thermocouple rever-
se polarity
Connect thermocoup-
le correctly
+ orange line
- white line
"Er05" displayed
Thermocouple short-
circuited or heater
defective
Check the heater/
thermocouple
and
have it/them replaced
if necessary.
"Er06" displayed
Defective
electronic
system
Check the electronic
system and have it
replaced if necessary.
"Er09" displayed
Power failure
If there is a brief
power failure during a
heating-up or cooling-
down phase and the
temperature is below
1000 °C, the process
will be continued once
power has been resto-
red. If the power
failure lasts longer
than 10 seconds and
the temperature is
above 1000 °C, the
process will be stop-
ped and the display
will flash.
The Start/Stopbutton
must be pressed to
acknowledge
error
Er09.
Содержание FINOSINT KOMPAKT
Страница 26: ...26 6 Save your port configuration in the File menu and Save As...
Страница 29: ...29 ENG 16Set the checkmark at Space 17Click on the Next button 18Click on the More button...
Страница 30: ...30 19Permute point and comma for both settings and confirm with OK 20Click on the Finish button and OK...
Страница 31: ...31 ENG 21The data series are displayed 22Select the Diagram Assistant and choose a diagram type e g line...
Страница 37: ...37 ENG...