Table 6 Keypad functions
Indicator lights and keys Description
ISO indicator light
Shows the last ISO code on any one of four sizes.
Code limit indicator lights
Shows the ISO, SAE and NAS code limits.
Alarm indicator light
Flashes along with the selected size indicator light when a cleanliness code alarm
limit is exceeded.
R SCROLL key
Moves down through the four sizes only when ISO, NAS or SAE is selected. Shows
any system or cleanliness code alarm. Flashes when the first key is pushed.
L SCROLL key
Moves down through the options on the left side of the display. Shows any system or
cleanliness code alarm. Flashes when the first key is pushed.
STATUS indicator light
Shows a numerical system status code. Flashes when there is a system alarm.
T °C indicator light
Shows the system temperature from the last completed sample (XX.X).
SAE indicator light
Shows the last SAE code on any one of the four sizes.
NAS indicator light
Shows the last NAS code on any one of the two differential groups (indicator lights
6–14 or 14–21).
Operation
Configure the instrument
1.
Open the
PC9001
Configuration Utility. A window opens (
).
2.
Select the communication (COM) port on the computer that is connected to the counter.
3.
Change the Node ID setting to the Node ID of the instrument. The default Node ID setting for new
instruments is 138. Display versions of the instrument show the Node ID for the first three
seconds after power up.
4.
Click the connect/disconnect button to connect to the counter.