Speed can be adjusted using the “
-
“
and “+” buttons. When moving away from this page the pump
will return to its normal speed.
ISO Alarm Limits
Alarm limits for low (clean) or high (dirty) can be set. Low refers to oil achieving required cleanliness
codes in all size categories. High refers to oil exceeding any one of the codes for particulate
contamination. When the oil is clean the ISO code on the main screen will turn green, when dirty it
will become red, between the two values the text is white.
Datalogging
The unit contains two internal dataloggers, one for the graph and one for the tests. Whenever the
unit is switched off the graph will be reset, but this can also be done from the screen without having
to switch off the unit. The main data log tied to the sensors will hold 100,000 lines of data (100,000
polls of the sensors) which equates to 138 days of testing.
To create a new log file the button needs to be held down for a couple of seconds, until a
confirmation message is displayed. This creates a new log file but does not delete the previous.
Clearing log files from the device must be done through the software, either the Particle Pal CSV
Reporting Tool or the Particle Pal Device Utility.