1.6.
Dimensions
1.7.
Technical Specifications
Type
Dust (D – black cable)
Hybrid (G/D – blue cable)
2994502301-C-S
2994502201-C-S
Sensing principle
Capacitive Ceramic
Detection range
± 300 mbarg
Overpressure
4 bar
Deflagration overpressure
12 bar
Vacuum resistance
Full vacuum
Process connection
G 1”
Wetted parts
1
Ceramic (Al
2
O
3
), 1.4404 (316 SST)
EPDM E7502 o-rings - FDA approved
Housing
1.4404 (316L SST), Aluminium
Temperature range
Process -20°C to +125°C
Cleaning Incidental +150°C
Ambient -20°C to +65°C
Storage -40°C to +80°C
Power supply
12 to 30 VDC
Current consumption
100 mA max
Output signal
4 – 20 mA current loop
Response time
2
less than 1 ms
Electrical connection
Fixed shielded and vented cable
4 x 0.22mm² Black or Blue - Ø 6.3mm (length 3m)
Approvals
CE
IP67
ATEX II 1/2 D Ex ta IIIC T85…T130°C Da/Db IP67
ATEX II 1/2 G Ex ia IIC T6…T4 Ga/Gb
ATEX II 1/2 D Ex ta IIIC T85…T130°C Da/Db IP67
EAC
EAC
IECEx INE 15.0001X
-
-
FM approval HLC 7/04
-
CSA 05 1668999/X
(1) Contact factory for version with FKM14-70.
(2) Response time definition: from pressure measurement to EPC controller output change.