3
3-15
Using a spectrophotometer
N
OTE
:
To view version and serial number information, click Show DTP-41 Info.
When finished, click Done.
10.
Remove the shipping spacer from the measurement page slot, if you have not already
done so.
11.
Insert the end of the Color Reflection Reference strip with the arrow into the
calibration strip entrance on the DTP41, centering it below the alignment mark
(see the diagram on page 3-8).
Insert the strip about 3 inches, past the front idler rollers, until it rests against the rear
drive rollers.
12.
Click Read Reference.
The DTP41 will pull the calibration strip through automatically.