22. Insert the lead and the LeadLoc into the lead holder making sure the tabs of the LeadLoc are aligned with the thumbknob
on the lead holder. Gently tighten the thumbknob to secure the LeadLoc into the holder then loosen slightly, ensuring
that the LeadLoc cannot move vertically.
23. While pinching the LeadLoc, move the lead so that the contacts are in the desired position relative to target depth.
Note- When using the 120mm inserter, extended clearance, the depth stop will be positioned at 319 + 30mm as shown in
the figure on page 9.
24. Release pressure on the LeadLoc to secure the lead. To confirm the lead does not move during the insertion process, use
an appropriate (sterile) pen to make a mark on the lead just above where it comes out of the LeadLoc.
25. Loosen the lead holder and remove from the measuring fixture.