Directions For Use

L011-61 (Rev F0, 2019-08-08)

Contains directions for the following products:
66-IT-MR1, 66-IT-MR3, 66-WP-ID, 66-WP-ID-01, 
66-WP-ID-02, 66-WP-MR

WayPoint™ Inserter System

FHC, Inc. 

1201 Main Street
Bowdoin, ME 04287 USA
Fax: +1-207-666-8292

24 hour technical service:

1-800-326-2905 (US & Can)
+1-207-666-8190

FHC Europe

(TERMOBIT PROD srl)
42A Barbu Vacarescu Str, 3rd Fl 
Bucharest 020281Sector 2
Romania 

FHC Latin America

Calle 6 Sur Cra 43 A-200
Edifi cio LUGO Ofi cina 1406
Medellín-Colombia

www.fh-co.com

Содержание Waypoint 66-IT-MR1

Страница 1: ...MR WayPoint Inserter System FHC Inc 1201 Main Street Bowdoin ME 04287 USA Fax 1 207 666 8292 24 hour technical service 1 800 326 2905 US Can 1 207 666 8190 FHC Europe TERMOBIT PROD srl 42A Barbu Vacar...

Страница 2: ......

Страница 3: ...Warnings and Cautions Specifications Inventory Storage Cleaning Sterilization Height Chart 120mm 130mm Inserter Height Chart 120mm Inserter extended clearance Pre Use Check Illustrative Procedure MRI...

Страница 4: ...re with FHC Inc 1201 Main Street Bowdoin Me 04287 USA Indicates the temperature limits to which the medical device can be safely exposed Indicates the range of humidity to which the medical device can...

Страница 5: ...Display module 66 EL MS Remote control 66 EL RM USB cord N5 55 02 Power cord country specific 1 2 3 4 11 10 8 7 13 6 12 66 WP MR includes 6 Burr Hole Marker 7 Standoff Wrench 8 Hex Wrench 9 Manual Ha...

Страница 6: ...od Container Manual Automated Tray fully loaded with parts in sonication unit with detergent Asepti Wash Plus liquid Reverse Osmosis de ionized water Use clean soft cloth Phase Recirculation Time in m...

Страница 7: ...ull cycle Gravity wrapped in 2 layers of 1 ply polypropylene wrap 3 Exposure Time 30 minutes at 132 C 270 F Minimum Dry Time 60 minutes 3 Cycle was validated using Halyard Health H300 wrap Prevacuum w...

Страница 8: ...L011 61 Page 8 of 25 2019 08 08 Rev F0 Shown with MR Lead Insertion Tube 66 IT MR1...

Страница 9: ...Page 9 of 25 L011 61 Rev F0 2019 08 08 Shown with MR Lead Insertion Tube 66 IT MR1...

Страница 10: ...d threads or looseness 4 Pull the stylet up 2cm out of the insertion tube Ensure that the stylet tip retracts into the insertion tube 5 Place the center hub into the integrated indexing ring Secure th...

Страница 11: ...thumbknobs until the drape support is held securely 7 Secure the center positioner to the inserter 8 Install the insertion tube with stylet into the inserter When the inserter and positioner are moun...

Страница 12: ...sertion tubes should always be carefully inspected before use because damage to them can cause targeting errors and impact patient safety Handle insertion tubes carefully to prevent bending 3 Install...

Страница 13: ...61 Rev F0 2019 08 08 4 Insert burr hole marker bushing into center hub 5 Insert the burr hole marking tool down through the bushing 6 Create indentation in skin skull for location of burr hole and re...

Страница 14: ...L011 61 Page 14 of 25 2019 08 08 Rev F0 7 Remove burr hole marker bushing 8 Remove platform and create burr hole 9 Reattach platform...

Страница 15: ...r or erratic function is observed discontinue use of the WayPoint Inserter System immediately and evaluate the potential impact to patient safety before continuing its unmitigated use WARNING Always c...

Страница 16: ...s in the unused indexing ring tapped holes and thread them in lightly 14 Mount the drape support as shown 15 Tighten the 8mm thumbknobs until the drape support is held securely The thumbknobs should b...

Страница 17: ...or center trajectory remains constant regardless of orientation Four parallel tracks at 2mm orthogonal offset are achieved with alignment to locations A C E and G designated as the black dots below a...

Страница 18: ...Loc to approximately 25 mm from the lead stylet handle making sure the lead is in the center notch of the LeadLoc 21 Release pressure on the LeadLoc WARNING Do NOT over pinch the tabs Over pinching th...

Страница 19: ...pinching the LeadLoc move the lead so that the contacts are in the desired position relative to target depth Note When using the 120mm inserter extended clearance the depth stop will be positioned at...

Страница 20: ...Remove the LeadLoc with lead from lead holder 27 Remove drape support by removing the two 8mm thumbknobs and then lifting the drape support straight up 28 Remove stylet Be sure to hold the collar of t...

Страница 21: ...is not being used the lead should be placed down through the selected track 31 Carefully secure the LeadLoc and lead in the lead holder making sure the tabs of the LeadLoc align with the thumbknob on...

Страница 22: ...ice like Medtronic Navigus cap 35 Remove the stylet from the lead 36 Loosen the lead holder thumbknob only slightly ensuring that it still retains the LeadLoc and that the LeadLoc cannot move vertical...

Страница 23: ...manual handle 41 Place the hex wrench over the first anchor to be removed Drop the combination driver down into the hex wrench until it engages with the anchor Twist the hex wrench counterclockwise to...