etatrack 360is issue 1
18
ALARMS
Facilities are provided for both low (B) and high (T) rate alarm settings. A master
OFF can be set for both during initial set-up of the FETATRACK 360. Either alarm
can be set on or off independently in this way.
In normal operation if alarm levels have been set, then the alarm can be enabled or
disabled by pressing both the volume control keys simultaneously. Alarm status is
shown on the display for a few seconds. Alarm settings are printed between the
chart scales on every page of the recording.
If an alarm condition arises the FETATRACK 360 emits an intermittent beep. The
chart recording indicates an alarm state with a down arrow marked with an “A” on
the top edge of the chart. The alarm is reset by pressing the either volume button.
A low heart rate alarm may be set between 90 and 120 BPM in steps of 5 BPM. A
high heart rate alarm may be set between 160 and 190 BPM in steps of 5 BPM. For
details of how to set the alarm levels, or to turn the alarms off, see the section
"SETTING UP THE FETATRACK 360".