THE PRINTER
The FETATRACK 360 prints on thermo-sensitive z-fold paper using a high
resolution thermal matrix printer. Heart rate, uterine activity and fetal movement are
recorded together with event marks, date, time, recorder speed and mode.
To load the paper pack in the printer, open the printer door by lifting the door
release lever and pulling the door open. The paper tray will slide forward
automatically, giving easy access for paper loading. Open two pages of paper from
the beginning of the paper pack and with the print uppermost and the Company
name and logo to the left place the pack in the bottom of the printer drawer. Make
sure that the paper is properly positioned; pull the free end of the pack over the
roller and close the door with the paper coming out over the top edge of the printer
door.
To operate the printer press the Printer On/Off switch on the front panel. The printer
will start to run and the chart speed will be shown in the display. To turn the printer
off press and hold the printer on/off switch until the printer stops. A few centimeters
of paper will be fed out at high speed.
The FETATRACK 360 printer will record at 1, 2 or 3 cm/min. To change the speed
during recording simply press the printer on/off switch momentarily. The
FETATRACK 360 is set to run at the speed of your choice and will always start at
this speed every time the printer is turned on. For details of how to change the
default speed see the section entitled "SETTING UP THE FETATRACK 360".
In the event that the paper pack is exhausted during recording the printer will
automatically turn off. The display will show "Paper Out" and the unit will emit an
audible bleep. Load a new paper pack as described above and close the recorder
door.
To recommence recording press the printer on/off switch momentarily. The printer
will print at high speed from its internal memory, that data gathered during the time
taken to change the pack. In this way a complete recording may be obtained without
any loss of information whilst the printer is reloaded. Once the memory printout is
complete the printer will continue to record at the selected speed.
The printer may be turned off while the paper pack is changed if the printer on/off
button is pressed and held before opening the printer door. In this case there is no
automatic printout of data at the recommencement of recording.