SYSTEM CHARACTERISTICS
MODULE 1
MODULE 2
MODULE 3
Function
Stimulator Main
Stimulator- Sensor
Sensor-Stimulator
Module
Lairdtech BT900
SPBT3.0DP2
SPBT3.0DP2
Frequency (MHz)
2402-2480
2402-2480
2402-2480
Max. declared output power (dBm)
8
2,6
2,6
Max. antenna gain (dBi)
+0,5
+0,5
+0,5
Max. E.I.R.P (dBm)
8,5
3,1
3,1
Max. E.I.R.P (mW)
7,08
2,04
2,04
The equipment must not be used next to or mounted above or below another piece of equipment. If this cannot be avoided, the
equipment should be checked for normal operation in the configuration in which it will be used.
The use of accessories other than those specified for the equipment is not recommended. This could result in an increase in emissions
or a decrease in the immunity of the equipment.
GUIDANCE AND MANUFACTURER’S DECLARATION– ELECTROMAGNETIC EMISSIONS
GUIDANCE AND MANUFACTURER'S DECLARATION - ELECTROMAGNETIC IMMUNITY
The
Fesia
Walk device is intended for use in the electromagnetic environment specified below. The customer or user of this device
should assure that it is used in such an environment.
Where
P
is the maximum output power rating of the transmitter in watts (W) according to the
transmitter manufacturer and d is the recommended separation distance in meters (m).
b
Field strengths from fixed RF transmitters, as determined by an electromagnetic site survey
a
,
should be less than the compliance level in each frequency range.
b
Interference may occur in the vicinity of the equipment marked with the following symbol:
EMISSIONS TEST
COMPLIANCE
ELECTROMAGNETIC ENVIRONMENT GUIDANCE
RF emmisions CISPR 11
Group 1
The
Fesia
Walk device uses RF energy only for its in-
ternal function. Therefore, its RF emissions are very
low and are not likely to cause any interference in
nearby electronic equipment.
RF emmisions CISPR 11
Harmonic emissions IEC 61000-3-2
Voltage fluctuations/Flickers emis
-
sions IEC 61000-3-3
Class B
Class A
Fulfill
The
Fesia
Walk device is suitable for use in esta
-
blishments other than domestic and those directly
connected to the public low-voltage power supply
network that supplies buildings used for domestic
purposes.
Immunity test
IEC 60601 test level
Compliance level
Electromagnetic environment
guidance
Electrostatic Discharge (ESD)
IEC 61000-4-2
±6 kV contact
±8 kV air
±6 kV contact
±8 kV air
Floors should be Wood, concrete or
ceramic tile. If floors are covered with
synthetic material, the relative humidi
-
ty should be at least 30%.
Electrical fast transient/bursts
IEC 61000 4-4
±2 kV for Power supply
lines
±1 kV for I/O lines
(input/output)
±2 kV for Power supply
lines
±1 kV for I/O lines (input/
output)
Mains power quality should be that of
a typical commercial and/or hospital
environment.
Surges
IEC 61000 4-5
±1 kV differential mode
±2 kV common mode
±1 kV differential mode
±2 kV common mode
Mains power quality should be that of
a typical commercial and/or hospital
environment.
Voltage dips, short interruptions and
voltage variations on power supply
input lines
IEC 61000 4-11
<5% Ut
( >95% dip in Ut)
for 0,5 cycles
<40% Ut
( >60% dip in Ut)
for 5 cycles
70% Ut
(30% dip en Ut)
for 25 cycles
>5% Ut
(>95% dip en Ut)
for 5 sec
<5% Ut
( >95% dip in Ut)
for 0,5 cycles
40% Ut
(60% dip en Ut)
for 5 cycles
70% Ut
(30% dip en Ut)
for 25 cycles
>5% Ut
(>95% dip en Ut)
for 5 sec
Mains power quality should be that of
a typical commercial and/or hospital
environment.
If the user of the
Fesia
Grasp device
requires continued operation during
power mains interruption, it is recom
-
mended that the be powered from and
uninterruptible power supply.
Frequency (50/60 Hz) magnetic field
IEC 61000-4-8
3 A/m
3 A/m
Power frequency magnetic fields
should be at levels characteristic of a
typical location in a typical commercial
or hospital environment.
Conducted RF
IEC 61000 4-6
3 vrms
150 kHz to 80 MHz
3 Vrms
d = 1,17 √P 150 kHz to 80 MHz
Radiated RF
IEC 61000 4-3
3 v/m
80 MHz to 2,5 GHz
3 V/m
d =1,17√P 80 MHz to 800 MHz
d=2,33√P 800 MHz to 2,5 GHz
TECHNICAL INFORMATION
05
05
TECHNICAL INFORMATION
INSTRUCTIONS FOR USE |
Fesia
Walk
53
52
INSTRUCTIONS FOR USE |
Fesia
Walk
EMI TABLES
Содержание F0X6-XM
Страница 1: ...INSTRUCTIONS FOR USE FESIA WALK FESIA FW01 IFU EN 2021 11...
Страница 4: ...GENERAL INFORMATION 01...
Страница 7: ...02 FESIA WALK DEVICE...
Страница 13: ...FESIA PRO APPLICATION 03...
Страница 22: ...TECHNICAL SUPPORT 04...
Страница 24: ...05 TECHNICAL INFORMATION...