User’s Manual S40-R
User’s Manual S40-R
FEMA ELECTRÓNICA - Page 9
Advanced Display (AD.DI) -
Segments linearization. The
instrument has 10 linearization segments (11 points). Each
linearization points is formed of a signal input value (“INPX”)
and an associated display value (“DISPX”). Enter the values
for each point from the Points menu (“PNTS”). It is required
that the input signal values “INPX” must increase for each
new point. When all points are entered, select “Act” value
to “On” in the activation menu. A check is automatically
performed when entering the points and when activating the
points. In case of error verify the data entered on the points.
Leave the menu pressing “LE” button. Reset option “REST”
deletes all existing points.
Password (PASS) -
Select a number to act as password. This
password will be requested when entering the Configuration
Menu. To deactivate the password select “Off”.
Factory Settings (FACT) -
Factory default configuration.
Select “yES” to activate the factory default configuration.
Version (VER) -
Firmware version installed.
Light (LIGH) -
Brightness. Select between 5 predefined
levels of brightness.
Menu options OPT1 and OPT2 give access to the configura
-
tion menus of the options installed at slots Opt1, Opt2 and
Opt3. This menu depends on the installed option. If there is
no option installed the instrument shows “
NONE
”. Control
modules R1 are controlled from the standard alarm menu
(see section 1.12.7).
Peak & Hold (P.HLD)
- Peak & Hold function detects sharp
drops in the displayed value, and holds the display if a display
drop is detected. To disable the hold function for a moment,
press any of the front buttons or wait for a predefined time :
Time 0
hold disabled (Off)
Time 1 a 3999
seconds, before disabling the hold
Time 4000
infinite hold
The counting of seconds is started each time there is an in-
crease in the display value. Alarms will follow the input signal
while hold is active.
1.10.7 Tools
1.10.8 Menu OptX - Options
When the instrument detects that the displayed value does
not correspond to the expected value, the display will flash
and alternate with a message.
“
h.udr
” Hardware underrange. The input signal is negative.
“
h.oVr
” Hardware overrange. The input signal is higher
than excitation voltage (+5V).
“
d.udr
” Display underrange. The instrument is displaying
the minimum value (-1999) and can not display below.
“
d.oVr
” Display overrange. The instrument is displaying the
maximum value (9999) and can not display above.
“
hoLd
” The instrument is showing the value present when
the hold function was activated. Hold function is active.
“
Min
”
The instrument displays the minimum displayed
value in memory. The minimum visualization is active.
“
MAX
” The instrument displays the maximum displayed
value in memory. The maximum visualization is active.
“
P.hLd
” The instrument displays the peak value. Function
Peak&Hold is active.
“
Err.0
”
Values introduced on the “ScAL” menu are not valid.
Default values are activated. The slope defined by the two
points is almost vertical (higher than 5000):
“
Err.1
”
Password incorrect.
“
Err.2
”
The instrument has detected an installed option but
was unable to communicate.
“
Err.3
”
Related to the edition of a linearization point. Re-
turns to the point for further edition. The value of “InPX” of
the edited point is lower than “InPX” of the previous point.
“
Err.4
”
Related to the activation (“Act”) of the lineariza-
tion segments (“Ad.dI”). Activation is not allowed. The value
“InPX” of one of the points is higher than “InPX” of the next
point.
“
Err.5
”
Related to the activation (“Act”) of the linearization
segments (“Ad.dI”). Activation is not allowed. The slope de-
fined by one of the segments is almost vertical.
“
Err.8
”
Excitation voltage overloaded. Excessive current
is being demanded to the excitation voltage. Potentiometer
nominal value is lower than 150 Ohms.
“
E.101
”
Option is installed but the type can not be recog-
nized.
(dhI-dLo) [Counts]
(IhI-ILo) [mA or Vdc]
< 5000
Slope condition for Err.0 and Err5
1.11 Messages and errors