Setup > RH Conversion
The Setup RH Conversion menu is used to correct for the
temperature sensor being inside the F-920 housing. This causes
the temperature sensor to reflect the internal temperature of
the F-920 and not the ambient temperature. The Relative
Humidity is calculated based on temperature, so if the sample
environment is preferred over the internal F-920 temperature,
enter the correct sample environment temperature and select
Yes to utilize it with your data.
Sample T(C): Temperature of sample environment in Celsius
Use Sample T: Yes will utilize the Sample T to calculate RH%
No will utilize the F-920 internal T to calculate
RH%
Device RH(%): Relative Humidity calculated from the F-920
internal T in %.
Sample RH(%): Relative Humidity calculated from the sample
Environment T in %.