Sensors
Carbon Dioxide (CO
₂)
SENSOR
Infrared sensor, pyroelectric detector
Nominal Range
0-100%
Full Scale Resolution
0.01% absolute
Accuracy
–
Trigger
Mode
±.5% absolute and ±3% of measured
value
Accuracy
–
Continuous
Mode
±.01% absolute and ±3% of measured
value
Lower Detection Limit
.01%
Sampling Time
10 seconds
Zero Interval
Weekly
Calibration Interval
12 Months
Lifetime
5 years
Oxygen (O₂)
Sensor
Electrochemical
Nominal Range
0-100%
Full Scale Resolution
0.1% absolute
Accuracy
–
Trigger
Mode
±0.3% absolute and ±2% of measured
value
Accuracy
–
Continuous
Mode
±0.1% absolute and ±2% of measured
value
Lower Detection Limit
0.1%
Sampling Time
10 seconds
Zero Interval
Weekly
Calibration Interval
6 months
Lifetime
2 years