F-920 Instruction Manual rev. 5/10/2017
1554 NE 3
rd
Ave, Camas, WA 98607, USA
11
Phone: (360) 833-8835
[email protected]
www.felixinstruments.com
Setup > Parameters
The parameters option allows the user to control the sample
volume from 10.0 mL up to 29.9 mL when using Trigger mode.
The default volume is 10 mL.
Setup > Set Zero
The set zero process sets a new baseline, or zero, for the
sensors:
CO2
O2
Set zero should take place weekly for both the Oxygen (O2)
sensor and the Carbon Dioxide (CO2) sensor to maintain
maximum performance. To set zero, first press the right arrow
to select the sensor. Wait for a beep to verify the set zero
process was completed. The set zero process requires the use of
100% nitrogen gas for O2, and an external conditioning
chamber full of Soda Lime to set zero for CO2. Setting zero is
important in establishing a daily zero baseline for the sensors.
Mode
Measure
Trigger
Connection
USB Storage