background image

[email protected]

Page 10

Microscope Operation:

Adjust the Aperture Diaphragm

The aperture diaphragm decides the numerical aperture (NA) of 

the illumination. If the NA of the illumination matches the NA of 

the objective, the best resolution is obtained with increased 

contrast and depth of field.

Adjusting the aperture diaphragm is the same as adjusting the 

field diaphragm.

Adjust the size of the aperture diaphragm for comfortable 

viewing.

Using the Simple Polarizing Kits

The simple polarizing kit includes the polarizer (1) and the 360˚ 

rotating analyzer (2).

When using the polarizer remove the color filter from the light 

path.

Dialing the turntable on the 360˚ rotating analyzer can change the 

orthogonal states of the polarized light.

When the turntable on the analyzer is set to “0” the analyzer and 

polarizer are in the orthogonal state.

Center the Light Filament

Adjust the filament to the center of the field of view for even 

illumination. Shift the ground glass slider (1) out of the optical 

path until it clicks.

Set the field diaphragm lever (4) and the aperture diaphragm lever 

(5) to the left. Place a white piece of paper on the stage. Rotate the 

nosepiece without an objective in it into the light path.

Adjust the halogen bulb horizontal adjustment screw (2) and the 

vertical adjustment screw (3) to center the light. Adjust the focus 

lock handle (6) to get a clear filament image, then center it in the 

field of view.

Pull the ground glass slider back into the light path and put a 

sample on the stage, adjusting the focus knob for a clear image.

Observe through the eyepieces and once a clear, focused image is 

obtained, lock the focus lock handle (6). Adjust the field 

diaphragm (4) and aperture (5) for comfortable observation.

 

 

 

 

 

Replacing the Fuse

Before replacing the fuse, make sure the power is switched to “O” 

off and unplug the microscope. Use a screwdriver to remove the 

fuse (1) from the fuse holder (2) and replace it with a new one.

Fuse: 250V, 3A

Содержание Mi40

Страница 1: ...Mi40 Microscope User s Manual for Inverted Metallurgical Microscopes info feinoptic com...

Страница 2: ...ents Trinocular Port Fixed Stage Eyepieces Metal Stage Halogen Bulb Horizontal Adjustment Screw Field Diaphragm Centering Screw Power Switch Fine Focus Coarse Focus Coarse Tension Adjustment Beam Spli...

Страница 3: ...omponents Carrying Handle Polarizer 360 Rotating Analyzer Trinocular Port Diopter Adjustment Metallurgical Objectives Rheostat Control Aperture Diaphragm Adjustment Field Diaphragm Adjustment Fine Foc...

Страница 4: ...info feinoptic com Page 3 Microscope Assembly Follow the numbered diagram below when assembling the microscope 1 2 3 4 5 6 7 8 9 10 11 12...

Страница 5: ...ugh room for any heat to dissipate Before replacing the fuse make sure the power switch 4 is in the O off position Voltage range of 100 240V is supported Make sure voltage is in this range Only use th...

Страница 6: ...the Objectives Rotate the coarse focus knob 1 until the nospiece is in it s lowest location Install the objectives 2 into the microscope nosepiece from the lowest magnification to the highest in a clo...

Страница 7: ...lot 3 on the illuminator connector panel It should click into place Reattach the handle 2 that was removed in order to slide the slider into the slot Assemble the Metal Stage Insert Insert the metal s...

Страница 8: ...Y Axis adjustment 6 in order to move the sample to the desired location Working range is 120mm x 80mm Connect the Power Supply Ensure the main switch is in the O off position Insert one side of the wi...

Страница 9: ...the sample on the stage and move the 5x objective into the light path The surface of the sample should be perpendicular to the objective Adjust the right diopter adjustment ring to 0 and look through...

Страница 10: ...gm The image of the field diaphragm can be seen through the eyetube Adjust the right and left centering screws 2 on the field diaphragm until the image is centered Open the field diaphragm slowly If t...

Страница 11: ...ust the filament to the center of the field of view for even illumination Shift the ground glass slider 1 out of the optical path until it clicks Set the field diaphragm lever 4 and the aperture diaph...

Страница 12: ...x Using the C Mount Adapter Loosen the set screw 1 on the trinocular port and remove the dust cover 2 Remove the dust covers from the c mount adapter 3 and place the c mount on the trinocular port tig...

Страница 13: ...rtical to the objective Adjust sample Aperture diaphragm not open Clean eyepiece objective Beam splitter is not in correct position Adjust beam splitter One side of field of view is dark or the image...

Страница 14: ...ck wire connections and connector pins for the bulb Bulb flickers or the brightness is not stable Coarse focusing knob is too tight Tension knob is too tight Loosen tension knob slightly Light does no...

Страница 15: ...lver screw on the light housing as shown at right The panel will open Remove the panel Depress the gold prongs as shown below and insert the bulb using a glove or lens tissue paper do not touch the bu...

Отзывы: