FZ12 Microscope Trouble Shooting

14

Fein 

TROUBLE

POSSIBLE CAUSE

SOLUTION

Bulb does not turn on.

Power cord is not plugged in.

Plug in power cord.

Power switch is not turned on.

Turn on power switch.

Bulb turns on, but field 

of view is dark.

Rheostat is too low.

Increase light intensity by 

rotating rheostat control.

Dust or dirt is visible in 

the field of view.

Dirt or dust is on the specimen.

Clean the sample.

Dirt or dust on the eyepieces.

Clean the eyepieces.

Dirt or dust on the stage.

Clean the stage.

Image has glare (hot 

spots).

Aperture diaphragm is closed down too much. Open the aperture diaphragm 

more.

Image is not sharp, 

contrast is poor and 

details are poorly visible.

Objective lens is dirty.

Clean objective lens.

Dirt or dust on eyepieces.

Clean eyepieces.

Field of view in one eye 

does not match that of 

the other.

Interpupillary distance has not been set.

Adjust interpupillary distance.

Incorrect diopter adjustment setting.

Adjust the diopter.

Coarse / fine adjustment 

knobs will not rotate 

easily or at all.

Tension adjustment ring is set too tight.

Adjust tension adjustment.

Image is not appearing 

through eyepieces or 

camera

Beam splitter is not in the proper position.

Check the position of the beam 

splitter.

Image is not clear on 

monitor when focusing 

microscope.

C-Mount adapter is not focused. 

Focus the C-Mount adapter.

Microscope body falls 

once focused and image 

falls out of focus.

Focusing knobs are too loose.

Tighten focusing knob.

[email protected]

Содержание FZ12

Страница 1: ...Fein FZ12 Microscope User s Manual info feinoptic com...

Страница 2: ...Fein FZ12 Microscope Components Beam Splitter Fine Focusing Knob Main Objective Aperture Diaphragm Adjusting Knob Coarse Focusing Knob BF DF Reflector Knob Zoom Knob Tension Adjustment Knob FZ12 BFDF...

Страница 3: ...2 Fein FZ12 Microscope Components Lock Knob Focus Holder Microscope Body Trinocular Head Eyepieces Stage Clips Stage Plate FZ12 BFDF...

Страница 4: ...2 Microscope Components Trinocular Head Eyepieces Focus Holder Fine Focusing Knob Coarse Focusing Knob Aperture Diaphragm Adjusting Knob LCD Display Intensity and Color Temperature Color Temperature C...

Страница 5: ...4 Fein FZ12 Microscope Components Zoom Knob Click Stop Rheostat Control Power On Off Main Objective FZ12 ILB Ultra Thin base Microscope Head...

Страница 6: ...FZ12 Microscope Assembly 5 Fein 3 5 6 7 1 2 8 9 4...

Страница 7: ...focus bracket group with a M4 inner hexagon Allen key Replace the decorative cover 1 Assemble the Objective Attach the objective 2 into the bottom of the microscope body 1 by rotating the objective un...

Страница 8: ...he eyepieces touches the bottom of the eyepiece tube Secure the eyepieces with the set screws 2 Assemble the Stage Plate Place the stage plate into the well space on the microscope stand To remove the...

Страница 9: ...rce Make sure the main switch is at the O off position Insert one end of the optical fiber 1 into the inner hole 2 of the power supply box aligning the flat surface with the lock screw 3 Secure the op...

Страница 10: ...d power off the light For a clear image ensure you do not leave fingerprints on the eyepieces or auxiliary lenses Wipe lenses gently with a soft tissue Wipe off fingerprints from lens surfaces with le...

Страница 11: ...ward and backward to achieve different lighting effects 1 2 1 Adjust the Focusing Tension If the handle is very heavy when coarse focusing or the specimen leaves the focus plane soon after focusing or...

Страница 12: ...djustment until it is clear Look through the left eyepiece If the image is not clear adjust the left diopter adjustment to get a clear image Left diopter adjusting ring Right diopter adjusting ring Ad...

Страница 13: ...se to disengage the detent function Do not over rotate the positioning screw as it will damage the hosing and internal mechanisms of the microscope Click Click Stop 1 Using the Eye cups The eyepieces...

Страница 14: ...o the trinocular tube and secure in place with the lock screw 1 Connect the camera to the c mount adapter Ensure the beam splitter pole 4 is in the outermost position for observation using the trinocu...

Страница 15: ...is poor and details are poorly visible Objective lens is dirty Clean objective lens Dirt or dust on eyepieces Clean eyepieces Field of view in one eye does not match that of the other Interpupillary...

Отзывы: