background image

FZ12 Microscope Operation:

10

Fein 

Set Illumination:

Plug in the LED light source and turn on the 

main power switch to “ON”.

Adjust the illumination intensity knob (2) 

until the illumination is comfortable for 

observation. Rotate the knob clockwise 

to increase the brightness and rotate 

counterclockwise to reduce the brightness.

Adjust the Reflector on BFDF Stand:

The reflector has two surfaces, a mirrored surface 

and a sandblasted aluminum surface. Both 

surfaces are reflective with the mirrored surface 

being more reflective.

Adjusting the reflector knob changes which 

reflective surface is in the light path. The reflector 

knob moves forward and backward to achieve 

different lighting effects.

1

2

1

Adjust the Focusing Tension:

If the handle is very heavy when coarse focusing 

or the specimen leaves the focus plane soon 

after focusing, or the stage declines itself, these 

problems can be solved by adjusting the tension 

adjustment ring (1).

Rotate the tension adjustment ring (1) according 

to the arrow direction in the figure, to tighten the 

focusing system; rotate the tension adjustment 

ring (1) in the opposite direction to loosen the 

focusing system.

1

2

Adjust Rheostat / Color Temperature on ILB Stand

The illumination intensity can be controlled by 

turning the Rheostat dial.

Adjust the color temperature by inserting the color 

temperature control tool and rotating to desired 

setting. LCD display will provide a read out of color 

temperature and illumination intensity.

1

2

Содержание FZ12

Страница 1: ...Fein FZ12 Microscope User s Manual info feinoptic com...

Страница 2: ...Fein FZ12 Microscope Components Beam Splitter Fine Focusing Knob Main Objective Aperture Diaphragm Adjusting Knob Coarse Focusing Knob BF DF Reflector Knob Zoom Knob Tension Adjustment Knob FZ12 BFDF...

Страница 3: ...2 Fein FZ12 Microscope Components Lock Knob Focus Holder Microscope Body Trinocular Head Eyepieces Stage Clips Stage Plate FZ12 BFDF...

Страница 4: ...2 Microscope Components Trinocular Head Eyepieces Focus Holder Fine Focusing Knob Coarse Focusing Knob Aperture Diaphragm Adjusting Knob LCD Display Intensity and Color Temperature Color Temperature C...

Страница 5: ...4 Fein FZ12 Microscope Components Zoom Knob Click Stop Rheostat Control Power On Off Main Objective FZ12 ILB Ultra Thin base Microscope Head...

Страница 6: ...FZ12 Microscope Assembly 5 Fein 3 5 6 7 1 2 8 9 4...

Страница 7: ...focus bracket group with a M4 inner hexagon Allen key Replace the decorative cover 1 Assemble the Objective Attach the objective 2 into the bottom of the microscope body 1 by rotating the objective un...

Страница 8: ...he eyepieces touches the bottom of the eyepiece tube Secure the eyepieces with the set screws 2 Assemble the Stage Plate Place the stage plate into the well space on the microscope stand To remove the...

Страница 9: ...rce Make sure the main switch is at the O off position Insert one end of the optical fiber 1 into the inner hole 2 of the power supply box aligning the flat surface with the lock screw 3 Secure the op...

Страница 10: ...d power off the light For a clear image ensure you do not leave fingerprints on the eyepieces or auxiliary lenses Wipe lenses gently with a soft tissue Wipe off fingerprints from lens surfaces with le...

Страница 11: ...ward and backward to achieve different lighting effects 1 2 1 Adjust the Focusing Tension If the handle is very heavy when coarse focusing or the specimen leaves the focus plane soon after focusing or...

Страница 12: ...djustment until it is clear Look through the left eyepiece If the image is not clear adjust the left diopter adjustment to get a clear image Left diopter adjusting ring Right diopter adjusting ring Ad...

Страница 13: ...se to disengage the detent function Do not over rotate the positioning screw as it will damage the hosing and internal mechanisms of the microscope Click Click Stop 1 Using the Eye cups The eyepieces...

Страница 14: ...o the trinocular tube and secure in place with the lock screw 1 Connect the camera to the c mount adapter Ensure the beam splitter pole 4 is in the outermost position for observation using the trinocu...

Страница 15: ...is poor and details are poorly visible Objective lens is dirty Clean objective lens Dirt or dust on eyepieces Clean eyepieces Field of view in one eye does not match that of the other Interpupillary...

Отзывы: