FZ12 Microscope Operation:
10
Fein
Set Illumination:
Plug in the LED light source and turn on the
main power switch to “ON”.
Adjust the illumination intensity knob (2)
until the illumination is comfortable for
observation. Rotate the knob clockwise
to increase the brightness and rotate
counterclockwise to reduce the brightness.
Adjust the Reflector on BFDF Stand:
The reflector has two surfaces, a mirrored surface
and a sandblasted aluminum surface. Both
surfaces are reflective with the mirrored surface
being more reflective.
Adjusting the reflector knob changes which
reflective surface is in the light path. The reflector
knob moves forward and backward to achieve
different lighting effects.
1
2
1
Adjust the Focusing Tension:
If the handle is very heavy when coarse focusing
or the specimen leaves the focus plane soon
after focusing, or the stage declines itself, these
problems can be solved by adjusting the tension
adjustment ring (1).
Rotate the tension adjustment ring (1) according
to the arrow direction in the figure, to tighten the
focusing system; rotate the tension adjustment
ring (1) in the opposite direction to loosen the
focusing system.
1
2
Adjust Rheostat / Color Temperature on ILB Stand
The illumination intensity can be controlled by
turning the Rheostat dial.
Adjust the color temperature by inserting the color
temperature control tool and rotating to desired
setting. LCD display will provide a read out of color
temperature and illumination intensity.
1