FEASA LED ANALYSER
ICT VERSION
Capture Mode
CAPTUREMULTI - Multiple Capture Command
Transmit
Receive
setrange@@#
putrange@@#
getrange@@
capturemulti
OK
OK
OK
OK
Where:
@@
represents the fiber number in the range 01 – 20.
#
represents the capture range 0 - 5
Description
The
capturemulti
or
cm
command will capture the LED data for the specified fiber at the
specified range.
The
setrange@@#
or
putrange@@#
commands will set the required range for each declared
individual fiber. The
getrange@@
will return the range the user has enables for this fiber.
Example:
The DUT has 6 leds with 5 very different intensity values. You can now specify fiber 1 to range
Ultra, Fiber 2 to range Super, Fiber 3 to range High, Fiber 4 to range Medium, Fiber 5 to range
Low and Fiber 6 to range Auto. Do one
capturemulti
or
CM
command to get your 6 readings.
Putrange015
(Fiber 1, Range Ultra / 5)
putrange024
(Fiber 2, Range Super / 4)
putrange033
(Fiber 3, Range High / 3)
putrange042
(Fiber 4, Range Medium / 2)
putrange051
(Fiber 5, Range Low / 1)
putrange060
(Fiber 6, Range Auto / 0)
capturemulti