Suction Unit 

 

F-31 

S e r v i c e   M a n u a l

       

 

 

Doc : Service Manual - F-31.doc Rev. 2 / 12 APR 2006 
Page 4 of 12 

 

BATTERY 

 

The units must be recharged when not in use.  

The appliance can be recharged with the appropriate external cable or by placing the appliance on the relevant 
recharge stirrup.  

BEFORE USING IT FOR THE FIRST TIME, IT IS NECESSARY TO KEEP BATTERIES UNDER CHARGE FOR 
24 HOURS NON-STOP. 

A battery that is constantly under charge lasts about 2 or 3 years.  

After this period of time it is recommended to replace it. 

If the appliance cannot be put under continuous recharge, recharge it for 24 hours every 15 days.  

When finished using it, the device must always be recharged.   

ATTENTION, THE BATTERY MAY BE DAMAGED IF NOT RECHARGED AFTER USE  

The battery that is completely flat and is not immediately recharged after use must be replaced by the authorized 
technical assistance. 

When recharging, always make sure the warning light on the front panel is on. 

1)  WHEN ON  THE DEVICE’S DISPLAY ONLY THE RED LED IS ON, WHILE THE GREEN LED AND THE TWO 

YELLOW LED ARE OFF, IT IS NECESSARY TO DISCONTINUE OPERATIONS AND RECHARGE THE DEVICE. 

2)  TO CHARGE THE DEVICE, KEEP TO THE FOLLOWING INSTRUCTIONS: CONNECT THE 230/110V PLUG ON 

THE BACK OF THE DEVICE, THE BLINKING LED SHALL TURN ON. SET THE SWITCH ON “CHARGE” AND 
KEEP CHARGING FOR AT LEAST 12 HOURS. 

THE DEVICE CAN OPERATE ON AN EXTERNAL 12 V BATTERY USING THE APPROPRIATE CABLE WITH 
CAR PLUG 

 
Battery change 
 

 

1)  Detach the connections (A) on the blue 

and brown wires 

 

2)  Unscrew the screws (B) 

 

3)  Raise the slab (C) 

 

4)  Replace the battery 

 

5)  Fix the slab (C) with the screws (B) 

 

6)  Connect the threads verifying the 

correspondence of the colors : 

 

Blue on positive pole 

Brown on negative pole 

 
 
 
 

Содержание F-31

Страница 1: ...Suction Unit F 31 S e r v i c e M a n u a l Doc Service Manual F 31 doc Rev 2 12 APR 2006 Page 1 of 12 NOTE This manual is related to the following Equipment Suction Pump F 31 ...

Страница 2: ...SING THE APPLIANCE UNDER DIFFERENT CONDITIONS THAN THOSE SPECIFIED IN THIS HANDBOOK MAY PREJUDICE ITS SAFETY AND ITS TECHNICAL PARAMETERS DELIVERY AND UNCRATING OF UNIT Inspect equipment and shipping crate immediately upon receipt If any damage is apparent you should both report it to the trucking delivery person and contact the transportation company immediately Make notes of any damage on the Bi...

Страница 3: ...WER SUPPLY 12V or 110 230V 50 60 Hz MAX VACUUM bar 0 70 MAX FLOW WITH FILTER l mins 10 MAX FLOW DIRECT MEASUREMENT ON ASPIRATION VENT l mins 18 BATTERY code 05 0040 12V 3 3AH FUSES A 2x1 25 ACCESSORIES INCLUDED CODE N DESCRIPTION 05 0001 OR 05 0002 JAR 1000 ML OR 2000 ML 05 00552 SET OF TUBES 05 0010 ASPIRATION PROBE CH 14 DISPOSABLE 05 0016 ANTIBACTERIAL HYDROPHOBIC FILTER DISPOSABLE ...

Страница 4: ... is completely flat and is not immediately recharged after use must be replaced by the authorized technical assistance When recharging always make sure the warning light on the front panel is on 1 WHEN ON THE DEVICE S DISPLAY ONLY THE RED LED IS ON WHILE THE GREEN LED AND THE TWO YELLOW LED ARE OFF IT IS NECESSARY TO DISCONTINUE OPERATIONS AND RECHARGE THE DEVICE 2 TO CHARGE THE DEVICE KEEP TO THE...

Страница 5: ...THE ASPIRATION 2 VENT 3 CONNECT THE OTHER END OF THE TUBE TO THE VACUUM 3 VENT WITH THE FLOAT ON THE JAR S LID 4 CONNECT THE LONG TUBE 4 TO THE OTHER PATIENT 5 VENT ON THE JAR S LID VENT WITHOUT FLOAT 5 CONNECT THE OTHER END OF THE LONG TUBE 4 TO THE PROBE CONNECTOR 6 AND EVENTUALLY THE SUPPLIED PROBE 7 PRESS THE ON OFF SWITCH 7 8 ASPIRATION INTENSITY CAN BE VARIED WITH THE RELEVANT CONTROL ON THE...

Страница 6: ... low battery may damage the appliance itself while continuously recharging it does not cause any damage It is recommended to always check the battery s recharge level LIGHT INDICATOR DESCRIPTION OF THE BATTERY LEVEL LED ON GREEN MAXIMUM CHARGE 2 LED ON YELLOW MEDIUM CHARGE 3 LED ON YELLOW LOW CHARGE 4 LED ON RED NEEDS RECHARGE 5 LED NEXT TO THE BATTERY S CHARGE LEVEL INDICATORS RED WHEN THIS LED I...

Страница 7: ...Y WARRANTY ON APPLIANCE THAT UPON INSPECTION BY THE TECHNICAL SERVICE SHOWS SIGNS OF TAMPERING THE OPERATOR MUST EXCLUSIVELY REPLACE THE SUPPLIED ACCESSORIES IF NECESSARY IN LINE WITH THE APPLIANCE S USE INTENSITY ASPIRATION TUBES SHOULD BE STERILIZED AFTER EACH USE IN CASE OF REPAIR THE USER SHOULD CALL FAZZINI S R L TRANSPORTATION COSTS MUST ALWAYS BE AGREED WITH FAZZINI S R L BEFORE SHIPPING FA...

Страница 8: ...rvice Manual F 31 doc Rev 2 12 APR 2006 Page 8 of 12 Legenda F 31 xx 1 Filter 2 Vacuum drive 3 Jar drive to the filter 4 Disposable tube to the patient 5 Jar drive to the patient 6 Suction catheter CH 14 7 Main switch 8 Vacuum regulator 9 Vacuum gauge ...

Страница 9: ...Suction Unit F 31 S e r v i c e M a n u a l Doc Service Manual F 31 doc Rev 2 12 APR 2006 Page 9 of 12 General assembly F 31 xx ...

Страница 10: ...Suction Unit F 31 S e r v i c e M a n u a l Doc Service Manual F 31 doc Rev 2 12 APR 2006 Page 10 of 12 Functional layout Main Wiring ...

Страница 11: ... e r v i c e M a n u a l Doc Service Manual F 31 doc Rev 2 12 APR 2006 Page 11 of 12 Pneumatic circuit 7 6 5 4 3 2 1 1 Catheter to the patient 2 Suction regulator 3 Jar 4 Non rebeathing valve 5 Filter 6 Pump Motor 7 Vent ...

Страница 12: ...raph With a multimeter check the voltage between the positive and negative pole if the voltage value is not around 10V dc replace the battery see battery paragraph Situation Possible Cause Solution a Vacuum regulator do not work properly Turn the manifold SP 21 b Internal tubing are not properly connected Open the equipment and verify internal tubing connection c Internal tubing are choked Open th...

Отзывы: