Correcting Malfunctions 
 

 

 

Situation 

 

Possible Cause 

Solution 

Disconnected or loose cord 

Check connections and reconnect unit 

Faulty wall outlet 

Use another wall outlet 

Fuses are damaged 

Check the fuses inside SP.38.02  

ª

 

If fuses are damaged check the electrical insulation 
on the electrical apparatus  

Main rocker switch is damaged   In ON position, check the presence of the light ; if no 

light is detected replace with a spare switch 

Equipment will 
not turn on 

Internal wiring connection are 
not completed  

Open the equipment and verify internal electrical wiring  
(see functional layout) 

 

Situation

 

 

Possible Cause

 

Solution

 

Vacuum regulator do not work 

properly 

Turn the manifold SP.21 

Internal tubing are not 
properly connected 

Open the equipment and verify internal tubing 
connection 

Internal tubing are choked 

Open the equipment and verify internal tubing bends 

2 Suction 

fault 

d Pump 

fault 

Open the equipment ,disconnect the tubing and check 
the pump 

 
 

 

 
 

 

 

 
 

 

 
 

 

 

 

 

 

 

 

 

 

 

Содержание F-18 Battery

Страница 1: ...n Pump F 18 F 36 F 20 F 40 S e r v i c e M a n u a l Doc Service Manual F 18 36 20 40 Rev 1 del 20 SEP 2005 Page 1 di 12 NOTE This manual is related to the following Equipment Suction Pump F 18 F 36 F 20 F 40 ...

Страница 2: ...bout the equipment spare parts and the information supplied CAUTIONS BEFORE STARTING TO INSPECT THE DEVICE MAKE SURE IT IS NOT CONNECTED TO THE ELECTRIC POWER SUPPLY NETWORK DO NOT USE THE APPLIANCE IN THE PRESENCE OF INFLAMMABLE AND OR EXPLOSIVE LIQUIDS GASES OR MIXTURES DO NOT LEAVE THE APPLIANCE EXPOSED TO ATMOSPHERIC AGENTS OR HEAT SOURCES USING THE APPLIANCE UNDER DIFFERENT CONDITIONS THAN TH...

Страница 3: ...XPOSED TO ATMOSPHERIC AGENTS OR HEAT SOURCES DO NOT LET CHILDREN OR INCAPABLE PERSONS USE THE APPLIANCE WITHOUT THE NECESSARY SUPERVISION BEFORE STARTING TO CLEAN THE DEVICE MAKE SURE IT IS NOT CONNECTED TO THE ELECTRIC POWER SUPPLY NETWORK DO NOT USE THE APPLIANCE FOR AIMS OTHER THAN THOSE FORESEEN BY THE MANUFACTURER USING THE APPLIANCE UNDER DIFFERENT CONDITIONS THAN THOSE SPECIFIED IN THIS HAN...

Страница 4: ...W VALVE WILL STOP ASPIRATION 11 IMMEDIATELY SWITCH OFF THE APPLIANCE BY SETTING THE ON OFF 7 SWITCH ON THE OFF POSITION THE MANUFACTURER ASSEMBLER INSTALLER OR IMPORTER ARE HELD LIABLE FOR THE SAFETY PERFORMANCE AND RELIABILITY OF THE APPLIANCE ONLY IF THE ELECTRIC SYSTEM TO WHICH THE APPLIANCE IS CONNECTED COMPLIES WITH L D 46 90 ACCESSORIES INCLUDED CODE N 05 0001 05 00552 05 0010 05 0015 DESCRI...

Страница 5: ... ON APPLIANCE THAT UPON INSPECTION BY THE TECHNICAL SERVICE SHOWS SIGNS OF TAMPERING THE OPERATOR MUST EXCLUSIVELY REPLACE THE SUPPLIED ACCESSORIES IF NECESSARY IN LINE WITH THE APPLIANCE S USE INTENSITY ASPIRATION TUBES SHOULD BE STERILIZED AFTER EACH USE IN CASE OF REPAIR THE USER SHOULD CALL FAZZINI S R L TRANSPORTATION COSTS MUST ALWAYS BE AGREED WITH FAZZINI S R L BEFORE SHIPPING FAILED PIECE...

Страница 6: ... Service Manual F 18 36 20 40 Rev 1 del 20 SEP 2005 Page 6 di 12 Legenda F 18 F 36 1 Filter 2 Vacuum drive 3 Jar drive to the filter 4 Silicone tube to the patient 5 Jar drive to the patient 6 Suction catheter CH 14 7 Main switch 8 Vacuum regulator 9 Vacuum gauge ...

Страница 7: ... F 36 F 20 F 40 S e r v i c e M a n u a l Doc Service Manual F 18 36 20 40 Rev 1 del 20 SEP 2005 Page 7 di 12 General assembly F 18 F 36 SP 32 05 Single head 230V motor for F 18 SP 32 01 Double head 230V motor for F 36 ...

Страница 8: ... Service Manual F 18 36 20 40 Rev 1 del 20 SEP 2005 Page 8 di 12 Legenda F 20 F 40 1 Filter 2 Vacuum drive 3 Jar drive to the filter 4 Silicone tube to the patient 5 Jar drive to the patient 6 Suction catheter CH 14 7 Main switch 8 Vacuum regulator 9 Vacuum gauge ...

Страница 9: ... F 36 F 20 F 40 S e r v i c e M a n u a l Doc Service Manual F 18 36 20 40 Rev 1 del 20 SEP 2005 Page 9 di 12 General assembly F 20 F 40 SP 32 05 Single head 230V motor for F 20 SP 32 01 Double head 230V motor for F 40 ...

Страница 10: ... M a n u a l Doc Service Manual F 18 36 20 40 Rev 1 del 20 SEP 2005 Page 10 di 12 Functional layout 4 2 3 1 Feeding wire Brown Feeding wire Blue Feeding wire Brown Feeding wire Blue 4 3 1 IEC Power inlets 2 M IEC double pole switch Feeding 50 60 Hz ...

Страница 11: ...0 S e r v i c e M a n u a l Doc Service Manual F 18 36 20 40 Rev 1 del 20 SEP 2005 Page 11 di 12 Pneumatic circuit 7 6 5 4 3 2 1 1 Catheter to the patient 2 Suction regulator 3 Jar 4 Non rebeathing valve 5 Filter 6 Pump Motor 7 Vent ...

Страница 12: ...er switch is damaged In ON position check the presence of the light if no light is detected replace with a spare switch 1 Equipment will not turn on e Internal wiring connection are not completed Open the equipment and verify internal electrical wiring see functional layout Situation Possible Cause Solution a Vacuum regulator do not work properly Turn the manifold SP 21 b Internal tubing are not p...

Отзывы: