Test of precision
Remove the tube from the manometer and hold the manometer in a
vertical position. When the pointer stands still at 0 on the scale, the
instrument has been adjusted properly. lf the pointer is below or above 0,
the instrument must be recalibrated. Send the instrument to your
specialized dealer.
Maintenance
In accordance with the testing requirements set down in the Medical Product
Law (MPG), a maintenance period of two years should be adhered to.
Technical Data
Minimal environmental conditions under which the error tolerance of +/-3
mm/Hg must be satisfied in accordance with the UNI-EN 1060/2 standard:
Measurement conditions:
10°C to 40°C at a relative air
humidity of 85 % (non-condensing).
Storage conditions:
-20°C to 70°C at a relative air
humidity of 85 % (non-condensing).
Versions:
Desk, wall, stand, and anaesthetic
models.
Type of indication:
Round scale.
Scale graduation
Increments of 2 mm/Hg.
Range shown on scale:
0 to 300 mm/Hg.
Measurement range:
0 to 300 mm/Hg.
Pointer movement
No stop pin.
Tube adapters:
“1 or 2” or “1 and 2” depending on
the particular model.
Pressure generation:
Bulb.
Pressure reduction:
Air-release valve that can be
regulated.
Subject to change without notice.
18