23
22
Setting Inputs / outputs
Current ouput 1
Contact output 1
Function selection: Counting / Alarm / Test On / Off
Counting :
Choice of flow variable (A / B / A+B pipe)
Counting direction ( / Reverse - / Net )
Pulse length
Pulse weight (value and unit)
Alarm :
Choice of the variable linked to the alarm
Contact output 2
Function selection: Counting / Alarm / Test On / Off
Counting :
Choice of flow variable (A / B / A+B pipe)
Counting direction ( / Reverse - / Net )
Pulse length
Pulse weight (value and unit)
Alarm :
Choice of the variable linked to the alarm
PT100 input (1-2-3-4)
Setting the characteristics of the Pt 100 input
•
Number of wires
•
Setting the temperature offset
Tests
Test mode of the analogue and contact outputs
Caution: in test mode, the output values will be changed.
Configuration of the internal recorder
Logger
File name
Test
File name
Define the name of the recording file associated with the measurement
point.
Number of data
Setting the number of variables to be recorded
Period
Choose a sampling period from a scrolling list
Data 1..2..3..
Define the nature of each variables
Pipe: Physical quantities (Flow, Speed, Speed...)
General: Device status, battery charge status
Type
Choose the type of value applied to all data
•
Average
•
A Min + Max
•
A Min + Max + Standard deviation
Menu
Measurement config.
Probe distance
US Signal display
Auto zero
Measurement config.
Measurement settings
Logger
Inputs / Outputs
Settings level
Logger
Settings
File management
Logger
Logger
Number of data
3
Logger
Period
10 s
Logger
Data 1
Pipe - A
Flow
Logger
Type
AVG, Min, Max
Menu
Measurement config.
Probe distance
US Signal display
Auto zero
Measurement config.
Measurement settings
Logger
Inputs / Outputs
Settings level
Inputs / Outputs
Current output 1
Contact output 1
Contact output 2
PT100 input 1
Inputs / Outputs
Inputs / Outputs
Inputs / Outputs
Inputs / Outputs
Inputs / Outputs
Inputs / Outputs
PT100 input 2
PT100 input 3
PT100 input 4
Tests
Current output 1
Contact output 1
Contact output 2
PT100 input 1
Current output 1
Contact output 1
Contact output 2
PT100 input 1
Current output 1
Contact output 1
Contact output 2
PT100 input 1
Current output 1
Contact output 1
Contact output 2
PT100 input 1
Adjustment of the characteristics of the 4-20 mA analogue output
Activation / deactivation /
•
Selection of the variable to be assigned to this output
•
Flow rate corresponding to 4 mA
•
Flow rate corresponding to 20 mA
•
Flow rate unit
•
Default value in case of fault (in mA)