Chapter 4: Spare parts and Maintenance
53
53
53
53
Disconnect carefully the concerned tranducer's cables.
Measure the transducers' capacitance.
The value must be about 2.9 nF.
•
If the value is ≈ 2 nF, one active element is faulty.
•
If the value is ≈ 1 nF, 2 active elements are faulty.
•
If the value is ≈ 0 nF, all active elements are faulty.
In all these cases, the transducer must be replaced.
Pull the cable out of the converter enclosure.