DazScope
manual
Chapter 3 : Software overview
algorithm from sending a dangerous waveform
into the Dazzler and the laser chain, the software
checks that every waveform to be loaded is safe, which means:
the waveform lies in the pulse shaping capability of the Dazzler.
the diffraction efficiency stays within a safe range.
Two parameters are involved in the security settings.
Acceptable Energy Loss Factor:
maximum fraction (0-1) of the acoustic energy that
can be ”out of the crystal”.
Max power:
maximum power level.
The user is strongly recommended to leave the default values unchanged.
3.3.4
Selecting and starting a measurement
To set a measurement from the loaded measurements list, select the corresponding item in the
Algorithm
listbox located in the control area and press
Run
. This updates the display area and
opens the related data processing software.
2
a truncated waveform for example.