![Faro SIDEREA Скачать руководство пользователя страница 30](http://html1.mh-extra.com/html/faro/siderea/siderea_users-manual-and-installation_3506679030.webp)
29
29
SID
èREA
7. PROBLEMS GUidE
The lamp does not
light.
The lamp does not
adjust.
The environment
lamp automatically
changes the lighting
mode.
The light emitted
from the
environment lamp
does not correspond
with the remote-
control scenarios.
Mains power is
missing.
The remote-control
battery is flat.
The distance between
the environment
lamp and the remote
control is too great.
The remote control
used is that of another
environment lamp.
The remote control
has been reset or has
lost the internal code.
The “Synchro”
function is active
with the dental lamp.
The remote-control
settings have been
modified.
Check tha mains power supply.
Check that the network voltage is
correct.
Change the battery.
Return the remote control within range.
Check with the other available remote
controls.
Carry out the pairing once more as
described in the user instructions section.
Disable the “Synchro” function using the
dental lamp control.
.
Carry out a reset of the scenarios using
the remote control as described in the
user instructions section.
User
User
User
User
User
User
User
User
Effect
Cause
Action
Resp
For all other problems, contact the Faro Spa authorized technical services.