20
Lampada dentale
AL
YA
the light does not switch on
– Check that the power supply is on.
– Check the state of the fuses.
– If none of these causes is verified, consult technical
assistance.
the light intensity is considerably
–
Clean the parabolas.
reduced
If the light intensity does not return to initial values,
consult technical assistance.
the handles do not hook on, or they
– Carry out a sterilisation cycle at 121º/134º with the specific
are difficult to hook onto the machine.
core former PN 310136.
Patches have appeared on the
– Clean the surfaces with the specific product “Faro Reflex”.
reflectors (dishes)
– Clean the surfaces with isopropyl alcohol.
– An unsuitable product for cleaning or disinfection has
been used that has damaged the surfaces.
– Contact customer care FARO.
Power voltage (without transformer)
:
17÷24V ac ± 10%
-
50/60 Hz
22÷35V dc
Power voltage (with transformer)
:
230 V 50/60 Hz
Absorbed power
:
26 W
Fuses (version with transformer)
:
2 x t250ma 250V
Protection against electrical hazards
:
Device class II
oPtICal SPECIFICatIonS In CoMPlIanCE WItH ISo 9680
Luminous spot dimensions
:
180 mm x 90 mm
Lux
:
3000*-50000* lux @700mm
Colour temperature
:
5000 K
CRI (color rendering index)
:
>95*
* typical value
TEcHnIcAL SPEcIFIcATIOnS
AcOUSTIc SIGnALS
Label compliant with EN 60825-1:94+A1:02+A2:01
TrOUBLESHOOTInG
OTP*
=
Beep 30 seconds
MAX
=
1 beep - 2 beep with proximity
MIN
=
1 beep
1 BEEP =
On the commands
1 BEEP =
At start-up
OpL**
=
Beep 30 seconds
* OTP: Over-temperature protection
** OpL: LED charger unplugged