Isolation Voltage
Port-to-earth ground
Continuous
60 V DC, Measurement Category I
Note
The TRC-8543 COM signals are not connected to the host port
ground.
Caution
Do not connect the product to signals or use for measurements
within Measurement Categories II, III, or IV.
Attention
Ne pas connecter le produit à des signaux dans les catégories
de mesure II, III ou IV et ne pas l'utiliser pour effectuer des mesures dans
ces catégories.
Warning
Do not connect the product to signals or use for measurements
within Measurement Categories II, III, or IV, or for measurements on MAINs
circuits or on circuits derived from Overvoltage Category II, III, or IV which
may have transient overvoltages above what the product can withstand.
The product must not be connected to circuits that have a maximum
voltage above the continuous working voltage, relative to earth or to
other channels, or this could damage and defeat the insulation. The
product can only withstand transients up to the transient overvoltage
rating without breakdown or damage to the insulation. An analysis of
the working voltages, loop impedances, temporary overvoltages, and
transient overvoltages in the system must be conducted prior to making
measurements.
Mise en garde
Ne pas connecter le produit à des signaux dans les
catégories de mesure II, III ou IV et ne pas l'utiliser pour des mesures dans
ces catégories, ou des mesures sur secteur ou sur des circuits dérivés de
© National Instruments
17
TRC-8543