Model 640 & 640CC Instruction Manual
D01206676
Revision A February 2016
22
10
Front Panel Controls
All the functions of the consistometer are controlled from the front panel. It is very
important for the user to have a thorough understanding of each control and its
effect on the operation of the consistometer.
The front panel controls can be roughly divided into four different sections: the
hydraulic pressure controls, the pneumatic controls, the cooling water controls, and
the electrical/electronic controls. This chapter will discuss each section in detail.
Hydraulic Pressure Controls
10.1
This section consists of the following controls: the PRESSURE gauge, the
PRESSURE RELEASE valve, and the FILL/DRAIN switch. Components that
make up this section are used to control the flow of oil used to pressurize the
cylinder and to display the cylinder pressure.
The pressure gauge displays pressure in both English and SI
units.
The pressure gauge displays pressure in both English and SI units.
The PRESSURE gauge is used to display the pressure inside the pressure vessel.
The PRESSURE RELEASE valve is used to release pressure from the pressurized
cylinder. The PRESSURE RELEASE valve must also be fully opened to remove
oil from the cylinder. If it is necessary to fill only one pressure vessel of a dual
consistometer, the PRESSURE RELEASE valve on the unfilled side must be
closed or oil may back up into the cylinder through the oil return lines.
The FILL/DRAIN switch is used to control the flow of oil into the pressure vessel.
When the switch is set to FILL and there is air pressure on the instrument, oil will
begin to flow into the cylinder. Ensure that the lid is closed and thermocouple
fitting is loose to allow air to escape while filling.
When the switch is set to DRAIN, the PRESSURE RELEASE valve is open, and
air is exhausted from the instrument, the cylinder will be emptied of oil. Under
DRAIN conditions, the cylinder must be closed. When the selector is in the OFF
position, the instrument is idle and cannot be run.