Model 35 Viscometer Instruction Manual
208878
Revision N, February 2013
7
2
Safety
Safe laboratory practices and procedures should be observed while operating and
maintaining the Model 35 viscometer.
The safe operation of the Fann Model 35 series viscometer requires that the
laboratory technician be familiar with the proper operating procedures and potential
hazards associated with the instrument.
The operator should be properly trained before operating this equipment. The safe
operation of this equipment may be impaired if it is used in a manner not specified
by the manufacturer.
2.1
Safe Electrical Operation
This instrument is driven by 115 volt or 230 volt electrical power. Keep hands,
clothes and other objects away from the rotating parts of the machine.
The optional heated sample cups and recirculating sample cups are electrically
heated. Make sure the power cord and other wiring associated with these cups and
the Model 35 assembly is in good condition and properly grounded.
Make sure the viscometer’s power switch is the OFF position and unplugged from
the source before cleaning, repairing or performing maintenance.
Please note that when motor speed /control power switch is in the
OFF position, power is only removed from the motor. However, the
power switch and the capacitor wil still have a charge.
Do not allow the viscometer base to get wet. If samples have been spilled or
splattered, wipe clean with a damp cloth. Do not allow water to run into the base;
excessive water could cause damage to the electrical components.