Section 3: Scale Installation
03/08
13
50612 Rev. 9
INSTALLATION, CONTINUED
7. Wire the scale cable to the proper type indicator, as shown in the chart below.
8. Once the scale platform is completely wired to the indicator, calibrate the unit.
─
Follow the appropriate indicator service manual to ensure a good calibration.
─
Typically, the Load Cells are pre-wired and connected.
Analog Interface (Junction Box 67171)
WIRE COLOR
FUNCTION
ANALOG INSTRUMENT
Black (
─
) Excitation
(
─
) Excitation
Green
(+) Excitation
(+) Excitation
Yellow Shield
Shield
White
(+) Signal
(+) Signal
Red (
─
)Signal (
─
)Signal
Note:
Models manufactured before
March 2008
may contain the following load
cell parts, and use the
West Coast wiring style
:
y
58902 – 1K, 2.5K
y
12896 – 5K
y
63593 – 10K
Wiring Connections for an Intalogix
™
Technology
Interface (QMB 15291)
QMB Terminal
WIRE COLOR
FUNCTION
ANALOG
INSTRUMENT
1 Green
(
─
) Excitation
1
2
Red
(+) Excitation
2
3 Black Ground
3
4
White
D Out
4
5 Brown D
In
5
6
Blue
EOC
6
7 Orange SCLK
7
8
Yellow
CS
8
9 Violet
Temperature 9
10
Gray/Shield
Chassis
10