1. 

Proper use and application 

1.1   Use 

Operating  table  SU-14  is  designed  for  supporting  the  patient  during  carrying  out  treatments  and 
operations.  The  table  with  the  proper  accessories  can  be  used  for  treatments  and  operations  in  the 
range  of  general  surgery,  vascular  surgery,  cardio  surgery,  neurosurgery,  urology,  proctology, 
gynecology,  laryngological,  ophthalmology,  endoscopies,  laparoscopy,  traumatic,  oncology,  dental, 
plastic surgery and others. Table top segments are X-ray permeable. 

1.2   General requirements 

The  product  is  intended  to  be  used  indoors.  Required  climatic  conditions:  temperature  from  +10  to 
+45ºC,  acceptable  change  of  surrounding  temperature  during  8  hours  should  not  exceed  20°C, 
relative humidity of the air should range from 30 to 80%, atmospheric pressure from 700 to 1060 hPa. 
The  product  should  be  used,  maintained  and  serviced  according  to  the  instructions  contained  in  this 
manual. 

 

 

Caution! 

Using,  maintaining  and  servicing  the  product  in  other  way  than  indicated  in  this 
manual is not permitted  and may lead to damages for which the user is to blame 
and for which the producer is not responsible. 
 

Installation  of  other  accessories  than  those  offered  by  the  producer  for  the 
product is allowed only on the basis of a written acceptance of the producer.  

1.3   Duties of the user  

User: any individual or corporate body who uses the product as its owner, lessee, pledge or who has 
a different right to the product as well as an entity who uses the product on its own or on whose behalf 
it is used.  

The user must ensure that the product shall be used exclusively in conformity with its destination and 
that it is used in appropriate conditions and in consistence with this manual. The user is also obliged to 
take  all  necessary  precautions  in  order  to  prevent  all  life  and  health  hazards  concerning  the  user, 
patients  and  any  third  party.  Only  authorised  persons  who  underwent  special  training  and  are 
acquainted with this manual may operate the product. The user must also ensure that all persons who 
operate the product have read, understood and apply instructions contained in this manual.  

1.4   Description of product 

The operating table consist of table base and table top that are permanently connection.  
In  standard  version  the  table  has  a  mobile  base  ,  equipped  with  central  blockade,  which  is  released 
with a foot lever. 
The  table-top  consists  of  segments,  which  number  and  configuration  has  constants  number  and 
configuration  (according  to  a  customer  special  order  this  number  will  be  change).  Segments  of  table 
top are X-ray permeable. The table top is provided with polyurethane mattresses. On both sides of the 
top there are side strips which allow installing accessories. 
Change  of  the  angle  of  table  top  segments  as  well  as  movement  of  the  column  are  adjusted  by 
mechanic  electro-mechanics  systems.  In  standard  version  the  electric  and  electromagnetic 
movements of table can be controlled by wire-connected controller. 
Thanks to the wide range of offered accessories the table can be used for different types of operations 
according to client wish. List of accessories you can find in Paragraph 4 - Installation and operation of 
accessories

The main power outlet is connected to the table by the power cord.  In order to disconnect, the user 
must unplug the power cord from the main power outlet. 
 

 

The  producer  reserves  the  right  to  introduce  in  the  product  structural  modifications  resulting 
from technical progress which are not covered in this user manual. 

The  producer  reserves  that  all  parameters  and  accessories  can  be  modified  or  change, 
especially  construction,  technology  and  materials,  not  lowering  accepted  parameter 
technically - user and safeties of products. 
 
 
 
 
 

Содержание SU-14

Страница 1: ...USER MANUAL Operating table SU 14 Factory number Edition 10 00 Issuance date March 2017 FAMED ŻYWIEC Spółka z o o Appendixes 2 3 List of Spare Parts ...

Страница 2: ...irements of the directive contained in Appendix 1 The compliance procedure was carried out in accordance with appendix VII of the directive Manufacturer Famed Żywiec Sp z o o Fabryczna 1 34 300 Żywiec POLAND Phone numbers Info line 48 33 866 63 75 24 hours a day Head office 48 33 866 62 00 Marketing Department 48 33 866 63 00 02 Export Department 48 33 866 62 50 52 Service 48 33 866 63 25 Fax 48 3...

Страница 3: ...use the guarantee for the product to be invalid Before starting any repairs the table must be disconnected from mains Operation table has been designed for a patient of 200kg of weight Notes concerned with safety The sign shown below says Warning Follow the instructions for safe use A label showing this sign is placed on any parts or mechanisms which may prove to be harmful to the patient or the p...

Страница 4: ...rg or reverse Trendelenburg and lateral tilt patient should be secured against uncontrolled sliding down from the operating table Use shoulder supports supporting rollers side supports belts and grips as securing elements Notes concerning start up operation and use The table during use should not be in the place which obstructs its disconnection from the mains When performing longitudinal Trendele...

Страница 5: ... the obligatory accessories for surgical tables Hand controller must be within reach of the medical staff Notes concerning disinfecting The product must not be disinfected in disinfecting chambers No bleaching agents containing active chlorine or oxygen caustic or corrosive chemicals are allowed No agents destroying the structure of plastic organic solvents can be applied to the plastic elements N...

Страница 6: ... CHANGE OF THE ANGULAR POSITION OF THE BACKRES SEGMENT 16 3 7 CHANGEOFANGULARPOSITIONOFTHEFOOTRESTS 17 3 8 INSERT AND TAKE OUT OF FOOTREST 17 3 9 INSERTING AND REMOVING FOOTREST 18 3 10 INSTALLATION AND REMOVALOF THE HEADREST 18 3 11 INSTALLATION AND DISASSEMBLY OF MATRESSES 18 3 12 T ABLE MOBILITY 18 3 13 ANTISTATICPROPERTIES 19 3 14 POTENTIAL EQUALIZING CLAMPS 19 3 15 COLISION 19 3 16 BATTERY CH...

Страница 7: ...o take all necessary precautions in order to prevent all life and health hazards concerning the user patients and any third party Only authorised persons who underwent special training and are acquainted with this manual may operate the product The user must also ensure that all persons who operate the product have read understood and apply instructions contained in this manual 1 4 Description of ...

Страница 8: ...User manual for Operating table SU 14 Page 8 1 5 Components and Functions description Fig 1 Operating table SU 14 27 ...

Страница 9: ... of the table top with terminal 650 mm 5 mm min table top height from floor the table stands on wheels 750 mm 15 mm max table top height from floor the table stands on wheels 1050mm 15 mm lifting angle of seat segment 85 lowering angle of seat segment 40 lifting angle of leg segment 20 25 lowering angle of foot support 90 Footrests widening angle 180 for two piece footrests angle of lateral tilt 2...

Страница 10: ...User manual for Operating table SU 14 Page 10 1 7 Use parameters The drawing below presents the main use parameter fig 2 ...

Страница 11: ...ts of the table or its equipment Throughout surgical procedures the table must be connected to the installation of potential equalization in the operations room When performing longitudinal Trendelenburg or reverse Trendelenburg and lateral tilting patient should be secured against uncontrolled sliding down from the operating table Use shoulder supports supporting rollers side supports belts and g...

Страница 12: ...ecommended distances between portable radio transmitters and the product Rated maximum output power of transmitter W 150 kHz to 80 MHz P d 2 1 distance in meters 150 kHz to 800 MHz P d 2 1 distance in meters 800 MHz to 2 5 GHz P d 3 2 distance in meters 0 01 0 1 0 1 0 2 0 1 0 4 0 4 0 7 1 1 2 1 2 2 3 10 4 4 7 100 12 12 23 For transmitters the maximum output power of which is not specified above the...

Страница 13: ...o a building where it is to be installed a product should be left for at least 12 hours Only after this time you can set a product in operation Laminar storage is permissible in accordance with the packaging marking In the absence of the marking storied storage is prohibited In case of the specific transport conditions particularly low temperature transport it is necessary to negotiate the way of ...

Страница 14: ...take the table from the palette and stop it add the headrest item 3 10 Installation and operation of the headrest place the product in its destination which meets the requirements described in this manual item 1 2 General requirements Connect the table to the mains voltage of 230V 50 60 Hz by connecting the power cord fig 1 Item 27 into the power socket at the base of the table and then to a wall ...

Страница 15: ...itch on through place the switching in I position Fig 1 pos 17 At this moment the table is activated and carrying out the movements through pushing the functioning control button is possible The buttons 1 to 7 are functioning buttons When the functioning button is pressed the movement represented by button symbol is operated so long as the button is pressed Caution The wire connected controller is...

Страница 16: ...d table top stops automatically Caution When performing longitudinal and side tilting functions patient should be secured against uncontrolled sliding down from the operating table Use shoulder supports supporting rollers side supports belts and grips as securing elements 3 5 Memory of table top position zero position The table controller enable writing 3 any table top position in this memory Ever...

Страница 17: ...olt 3 4 turns fig 1 item 23 and turning the footrest of the desired angle to the side Caution When the function is performed the footrests may collide with the base Check the proper clamp of the handle that locked the footrest turning 3 8 Insert and take out of footrest Footrest plate fig 5 item 1 is fixed by wedges fig 1 item 10 placed on the bottom of the seat segment by the blockages placed on ...

Страница 18: ...hen the headrest is put on the stripe one should determine its position by screwing down the screw fig 1 item 8 The headrest is dismounted by unscrewing of the screw and removal of the headrest from the stripe The angle of the headrest is changed similarly to the backrest segment that is by the handle which moves the mechanism of the gas spring When the lever is pressed fig 1 item 6 the required a...

Страница 19: ...ut surgical procedures the table must be connected to the installation of potential equalization in the operations room The cable is provided in standard table accessories 3 15 Collisions In some extreme positions of the table in particular when accessories installed on side stripes are used mechanical collisions may take place Because of that one should protect the table and accessories from dama...

Страница 20: ...ls WS 32 5 23 kg 18 Side x ray handle WS 33 0 14 kg 19 Wrist grip WS 34 5 14 kg 20 Attachment do arthroscopy WS 39 5 17 kg 21 Attachment for meniscus operation WS 40 5 17 kg 22 Belly belts WS 41 0 125 kg 23 Footrest belts WS 42 0 23 kg 24 Hand belt WS 43 0 17 kg 25 Thigh belt WS 44 0 23 kg 26 Support for hand surgery WS 48 5 17 kg 27 Womb support WS 49 5 40 kg 28 Side support WS 50 5 40 kg 29 Side...

Страница 21: ...after locking the segment should not change its position check by manual pressing of the segment 3 Check if there are no mechanical plays by manually moving the top of the table 4 Check the work of the electro mechanics system by performance of all movements controlled by wire connected control If the table underwent the above described tests with positive results and there were no disturbing soun...

Страница 22: ...ements concerning cleaning and disinfecting shall result in losing the guarantee for the product 6 3 Damages and defects Damages and defects found in the product or product accessories should be reported immediately to a person in charge of such issues The bed which can not be safely operated e g damaged electric or mechanical elements must not be used till it is repaired 6 4 Repairs and conservat...

Страница 23: ...mpliance check General technical condition check Table 2 Checking safety and functions check of functioning of remote operated functionalities check of reset function calibration of inclinometers checking collision limits checking antistatic parameters of mattresses Resistance value shall be within the range from 10 4 to 10 6 Q Check of electric connections Check of base and column condition Check...

Страница 24: ...spection should involve a visual inspection and the noticed malfunctions should be handled as provided in item 6 6 Location of defects and their removal 6 6 Location of defects and their removal damage Possible cause removal The table cannot be moved locked wheels Unlock blockade pedal Batteries discharged Charge the batteries Control plug incorrectly plugged to the socket in the table Plug in the...

Страница 25: ...n such an environment Immunity test IEC 60601 1 2 Test level Compliance level Electromagnetic environment guidance Electrostatic discharge ESD IEC 61000 4 2 6 kV contact 8 kV air 6 kV contact 8 kV air In the location of SU 14 use the floor should be wooden concrete or covered with ceramic tiles If the floor is covered with a synthetic material the relative humidity should be at least 30 Surge IEC ...

Страница 26: ...ecommended distance in meters m Field strengths from fixed RF transmitters as determined by an electromagnetic site survey a should be less than the compliance in each frequency range Interference may occur in the vicinity of equipment marked with following symbol a Field strength from fixed transmitters such as base stations for radio cellular cordless telephones and land mobile radio amateur rad...

Страница 27: ...particular boxes of a name plate 1 Name logo and address of the Manufacturer 2 Product symbol 3 General markers CE mark Warning Follow the instructions for safe use Caution During disposal of the product one should act according to the Act concerning electrical and electronic equipment waste 4 Product trade number with untypical order number if applicable 5 Index of the product 6 Class of duty 7 P...

Страница 28: ...User manual for Operating table SU 14 Page 28 8 2 Labels ...

Страница 29: ...oint see point 2 1 3 The method of battery charging 4 Warning Follow the instructions for safe use 5 For clearing and disinfection you should not use agents containing active oxygen and chlorine 6 Maximal workload 200 kg 7 Central wheels blockade 8 Back rest adjustment ...

Страница 30: ...nt 10 Foot rest adjustment 11 Hook for alimentation cable 12 Follow the instructions 13 Power socket Notice The producer reserves the right to introduce in the product structural modifications resulting from technical progress which are not covered in this operating manual ...

Страница 31: ... 1 C183701000005 2 Famed Żywiec Sp z o o 14 Earthing screw 1 C028001010088000 2 Famed Żywiec Sp z o o 15 Base cover 1 C183600000016 2 Famed Żywiec Sp z o o 16 Lock pedal left 2 S06549065006020 2 Famed Żywiec Sp z o o 17 Lock pedal right 2 S06549065006021 2 Famed Żywiec Sp z o o 18 Back rest mattress 1 C183702000003 2 Famed Żywiec Sp z o o 19 Servo motor for Trendelenburg 1 S09164004005711 2 Famed ...

Страница 32: ...nual for Operating Table SU 12 Page 32 Legend Markers in column Category 1 The elements endanger fast wear out 2 The elements not endanger fast wear out 3 The elements wear out depended of condition of exploitation ...

Отзывы: