Page 22
6 Technical operations
6.1 Storing
If the product is not to be used for a longer period of time, it should be stored in the below
mentioned climatic conditions:
•
temperature: 25º ±10ºC,
•
relative humidity: 50% ± 25%.
Caution!
During storing, the operating table must be turned off. The switch (position 17,
figure 1) must be in the “0” position.
If the table is stored for a longer period of time, it must be connected to the power
supply for 24 hours each 6 months to charge the batteries.
The table may not be stored with batteries low
6.2
Cleaning and disinfecting
For cleaning and disinfecting use cleaners, disinfectants recommended by Famed
Ż
ywiec Sp. z
o.o. in Annex 2 to this manual (allowed to turnover and use on the territory of the country, where
they are used).
After disinfection the Product should be washed with water to eliminate any stains.
After disinfecting dry thoroughly.
Dry with hot air (maximal temperature 60
°
C) or by wiping with a soft sterile cloth.
Caution!
Before disinfecting disconnect the product from mains!
The product/ table must not be disinfected in disinfecting chambers.
Bleaching (containing active chlorine or oxygen), caustic and corrosive agents
must not be used.
No disinfecting agents containing alcohol may be applied to elements made of
polyurethane (mattresses)!
No agents destroying the structure of plastic (organic solvents) may be applied to
the plastic elements.
Disregarding the above requirements concerning cleaning and disinfecting shall
result in losing the guarantee for the product
.
6.3
Damages and defects
Damages and defects found in the product or product accessories should be reported immediately
to a person in charge of such issues. The bed which can not be safely operated (e.g. damaged
electric or mechanical elements ) must not be used till it is repaired.
6.4
Repairs and conservation
Repairs are done by the producer. The user can not carry out any repairs on his own unless he
has undergone special training or has been authorised to do that. When the producer has given
his written permission for repair of the product by client’s technical staff, the producer shall provide
the client with necessary charts, lists of spare parts, descriptions and information on repairs.
The producer allows only using original spare parts. In order to provide safe and reliable operation
of the product one should use only spare parts provided by the producer. Worn out parts shall be
removed as provided in environmental protection regulations
Caution!
The product contains products which may be dangerous to the environment:
-
a battery integrated with controller – type BA 1812V 1,2Ah (2pieces).
The rules of proceeding with used products which may be dangerous to the
environment are defined in environmental regulations.
If there is a need to replace the batteries, used batteries are collected by the
medical device's producer or an authorised service point indicated by the
producer.
Used batteries receives a medical device manufacturer or indicated by the
authorized repairer or Third Party having required by environmental law
decisions to conduct.