2

 

The  WMA-005  and  005LF  amplifiers 
contain an over-voltage protection for daily 
use  in  the  laboratory  and  are  long-term 
short-circuit  proof.  The  input  (indicated 
“Input”  on  the  label),  output  (indicated 
“Output”) and DC power supply (Indicated 
“24V  DC  180mA”)  connections  are  all 
standard BNC. 
 
The  amplification  is  20x  (fixed).  The 
amplifiers have an output voltage range of 
–75V to +75V. The 24V DC power supply 
requirement  offers  the  possibility  for 
battery powered operation. With an output 
noise  below  3mVrms,  this  amplifier  will 
enable  many  measurements  in  different 
applications.  

 

 

Safety 

 
- This product is able to produce over 75V, 
but  the  50kΩ  output  impedance  limits  the 
current  to  a  rather  safe  maximum  of 
2.5mA. This voltage can be painful, but will 
not  be  lethal  (under  normal  circum-
stances). Nevertheless take care. This is a 
high  voltage  unit  and  safety  measures 
should be taken accordingly. 
-  The  internal  circuitry  of  the  amplifier 
operates  at  high  voltage.  Only  qualified 
personnel  from  Falco  Systems  should 
service it. 
- When the amplifier is turned on or off, a 
short  voltage  spike  may  appear  at  the 
output which may damage circuitry already 
connected to it. 
-  Replace  the  fuse  with  250V  315mAT 
5x20mm rated fuses only. 
-  This  product  is  only  suitable  for  indoor 
use  in  a  class  II  environment  (domestic, 
light industrial).   
-  The  power  supply  connected  to  the 
WMA-005/005LF 

should 

be 

double 

insulated (class II) or a SELV type (SELV 
=  Safety  Extra  Low  Voltage,  typically  a 
battery). 
-  The  WMA-005/005LF  is  a  class  III 
product.  Do  not  connect  the  amplifier  to 
the mains safety earth. If this is done, the 
system  becomes  a  class  I  apparatus  with 
class  III  isolation,  which  means  that  the 

protective  insulation  may  be  inadequate. 
Doing 

so 

(only 

under 

laboratory 

conditions)  is  fully  at  the  users’  risk.  Risk 
of electric shock! 
- The WMA-005/005LF should not be used 
above 2000m altitude to prevent electrical 
breakdown. 
- This product should only be cleaned with 
a  soft,  slightly  moist  cloth.  Unplug  the 
WMA-005/005LF  from  the  DC  power 
before cleaning. 
 
 

Input protection

 

 
A low-noise amplifier like the WMA-005 or 
WMA-005LF  can  never  be  made  fully 
insensitive to input overload conditions, as 
this  would  limit  the  performance  of  the 
amplifier to an unacceptably low level. The 
absolute  maximum  input  voltage  is  ±15V. 
Warning:  exceeding  this  spec  may  cause 
permanent malfunction of the amplifier and 
is not covered by the warranty! 
For  normal  operation,  input  voltages 
should  remain  in  the  -3.75V  to  +3.75V 
range,  resulting  with  an  amplification  of 
20x  in  an  output  voltage  swing  of  more 
than -75V to +75V.  

 

Ouput protection

 

 
The  WMA-005/005LF  has  been  designed 
to be fully stable with all capacitive loads. 
It  has  been  optimized  for  its  step-
response,  but  is  also  a  very  good  linear 
and  sine-wave  amplifier.  No  significant 
overshoot occurs at any capacitive load. 
Overloading 

or 

short-circuiting 

this 

amplifier will not break down the amplifier, 
due  to  the  50kΩ  current  limiting  resistor 
that has been employed at the output. 
 
Although the amplifier cannot be damaged 
by  a  short-circuit  condition  or  capacitive 
loading, two situations should be avoided: 
- Connecting a charged capacitor. 
- Using high inductance values (coils). 
  
 

 
 
 
 

 

Содержание WMA-005

Страница 1: ... e g for use in a Faraday cage Its internal high voltage generator is inductor free and hence does not spread magnetic interference in sensitive experimental setups Its large voltage range makes it an excellent choice for use with MEMS devices EO modulators piezo actuator positioning systems beam steering components and many more The amplifiers full power frequency response has a 3dB corner freque...

Страница 2: ...class III product Do not connect the amplifier to the mains safety earth If this is done the system becomes a class I apparatus with class III isolation which means that the protective insulation may be inadequate Doing so only under laboratory conditions is fully at the users risk Risk of electric shock The WMA 005 005LF should not be used above 2000m altitude to prevent electrical breakdown This...

Страница 3: ...e connection can cause overshoot in the oscilloscope reading The amplifier is capable of amplifying above 9kHz and should not be used for telecommunication as described in the R TTE directive 95 5 EC To prevent unwanted radiation all cables should be kept shorter than 30m Noise and offset The WMA 005 is a low noise amplifier The output noise is typically below 3mV rms If this is not low enough the...

Страница 4: ...tage Figure 3 Frequency response at 200mVpp output voltage 140 120 100 80 60 40 20 0 10 0 10 1 10 2 10 3 10 4 10 5 WMA 005 WMA 005LF Amplitude Vpp Frequency Hz 200 150 100 50 0 10 0 10 1 10 2 10 3 10 4 10 5 WMA 005 WMA 005LF Amplitude mVpp Frequency Hz ...

Страница 5: ...5 Figure 4 150Vpp 1kHz square wave Figure 5 Full scale 10kHz square wave Figure 6 30mVpp 1kHz small square wave output signal ...

Страница 6: ...6 0pF 470pF 2 2n 100pF 1nF 10nF Figure 7 Full power 1kHz square wave response with different capacitive loading conditions The 50kΩ output resistor limits the speed at which the capacitor can be charged ...

Страница 7: ...0 Noise less than 10mVpp or equivalently less than 3mVrms and typical DC offset 1 1mV for the WMA 005 The high voltage generator frequency around 150kHz contributes considerable to the overall noise This measurement was performed with 1m of coaxial cable connected to the output ...

Страница 8: ...h typical for the WMA 005 1 1mV offset typical Noise 0 1mVrms noise over the full bandwidth typical for the WMA 005LF 1 5mV offset typical Input impedance 100kΩ Output impedance 50kΩ Stability stable with all capacitive and resistive loads no overshoot 5 Fuse 1x 315mAT 230V Operating temperature 10 40 C Storage temperature 0 60 C Relative humidity 10 90 non condensing Maximum usage height 2000m Di...

Страница 9: ...alfunction of the product resulting from use or operation in other ways than specified in the user manual Malfunctioning due to misuse or abuse of the product Malfunctioning occurring after changes or repairs have been made by anyone other than Falco Systems To obtain warranty service inform Falco Systems via info falco systems com You will receive support and further instructions Falco Systems wi...

Отзывы: