Sterilisation and reuse can impair the function and is therefore not permissible either.
The LARYVOX
®
STYLE HME HighFlow filter cassette must under no circumstances
be inserted directly into the tracheostoma!
LARYVOX
®
STYLE TOP / LARYVOX
®
STYLE CAP
LARYVOX
®
STYLE TOP / LARYVOX
®
STYLE CAP are single-patient products and
are thus intended exclusively for one-time or repeated use in one individual patient.
Do not re-use with another patient. Do not reprocess for re-use by another patient.
Only clean LARYVOX
®
STYLE TOP / LARYVOX
®
STYLE CAP with cleaning agents
approved by the manufacturer.
LARYVOX
®
STYLE TOP / LARYVOX
®
STYLE CAP cannot be sterilised.
Heating to over 65 °C, disinfection with boiling water or steam sterilisation are not
permitted and lead to damage of LARYVOX
®
STYLE TOP / LARYVOX
®
STYLE
CAP.
LARYVOX
®
STYLE TOP / LARYVOX
®
STYLE CAP must never under any
circumstances be fitted directly into the tracheostoma!
V. COMPLICATIONS
If the particle filter is contaminated or if signs of increased respiratory resistance
appear, LARYVOX
®
STYLE must be removed immediately.
VI. PRECAUTIONS
Care must be taken to ensure that the valve opening is not blocked by clothing or the
like in order to enable unhindered breathing (risk of respiratory distress).
VII. CONTRAINDICATIONS
LARYVOX
®
STYLE should be removed if respiratory resistance is too high.
Use in patients with strong exsiccosis (dehydration), or excessive secretion in the
lungs and airways, or strong air leakage flow (exhaled air not flowing through the
filter cassette) is not permissible.
Not for use in patients with impaired consciousness and/or in persons unable to
look after themselves and requiring the help of others, who are not able to remove
Laryvox
®
STYLE by themselves in emergency situations.
VIII. INSTRUCTIONS
Check whether the packaging and the product are in a faultless condition.
Check the use-by or expiry date. Do not use after this date.
Place the LARYVOX
®
STYLE HME HighFlow filter cassette with the open side
facing upward (with the cross-pieces facing downward) on a firm, even surface ⑥.
Place LARYVOX
®
STYLE TOP with the reverse side on the LARYVOX
®
STYLE
HME HighFlow filter cassette ⑦.
Now press LARYVOX
®
STYLE TOP onto the LARYVOX
®
STYLE HME HighFlow
filter cassette until it clicks audibly into place ⑧.
Connect LARYVOX
®
STYLE TOP to the reverse side of LARYVOX
®
STYLE CAP
until it snaps into place ⑨.
Fit the LARYVOX
®
STYLE HME HighFlow filter cassette vertically onto the 22-mm
connection of the tracheostomy tube or the base plate, exerting gentle pressure ⑩.
In order to be able to use LARYVOX
®
STYLE with a 15-mm connector as well, use
LARYVOX
®
Adapter (REF 48920).
Check the respiratory resistance.
Remove LARYVOX
®
STYLE after a wearing time of not more than 24 hours ⑪, or if
the filter cassette is contaminated with secretions for instance.
To do this, hold the base plate or the tracheostomy tube steady with two fingers and
remove LARYVOX
®
STYLE by tilting it slightly downward ⑫.