![FagronLab BASIC Скачать руководство пользователя страница 35](http://html.mh-extra.com/html/fagronlab/basic/basic_user-manual_538547035.webp)
FagronLab™ BASIC – User Manual | 35
FagronLab™ BASIC
Notes on safety
•
FagronLab™ devices must only be connected to grounding type receptacles with 230 V/50 Hz(cps); 120 V/60 Hz(cps) or rated
country specific voltage installed according to the regulations of DIN VDE 0100
•
FagronLab™ devices have been designed for operation under normal room atmospheric conditions. Recommended values:
Ambient temperature between 15 - 30 °C/ 59 - 86 °F and relative air humidity less than 80 %
•
The device should be allowed to acclimatize for approx. 30 minutes at initial commissioning and/or after extended storage time
in cold rooms
•
The FagronLab™ BASIC should be placed to ensure easy access to the power switch and power cable also to avoid use by
unauthorized persons
•
Do not immerse FagronLab™ devices in water
•
Electronic parts should just be dis- and assembled by a certified service partner
•
Only operate the FagronLab™ stirrer inside a closed FagronLab™ jar
•
Do not touch rotating parts
•
Keep long hair away from rotating parts
•
During operation of the FagronLab™ BASIC: always keep long hair, body parts or objects away from the lifting mechanism.
Immediately turn off the power switch in case of an emergency or pull the power plug
•
Using the FagronLab™ devices not according to these operating instructions or with line products that the manufacturer did not
deliver or recommend may impair safety
•
FagronLab™ devices have not been designed for operation under hazardous conditions. Heed the relevant safety regulations
when handling hazardous substances (e.g. combustible liquids such as alcohol or similar substances)
•
FagronLab™ devices correspond to the safety standards for laboratory equipment. They have to be positioned to prevent any
interference or use by unauthorized persons
•
The device must not be disposed of in ordinary domestic waste. Please deliver the device to the available collecting and recycling
systems at the end of its life cycle
Technical data of the FagronLab™ BASIC
Electrical requirement
100 - 115 V / 220 - 240 V
Total power consumption
80 W
Power consumption (mixing motor)
80 W
Operating mode
continous operation S1
Safety class
I
Type of protection
IP 20
Connection
1 RS232 port
Rotational speed controller
in 10 steps electronic controlled
Timer
continuous electronically
FagronLab™ jar sizes
15-20 ml - 500 ml
Weight
22 lbs/10 kg
Dimensions (LxWxH in mm)
255x211x512
Testing certifications
CE
Tab: Technical data of the FagronLab™ BASIC