Section 7 |
Maintenance
The system does not require any daily maintenance. Avoid contact with any aggressive solvent like
pure alcohol, detergents containing alcohol (window cleaner), abrasive cleaning powders, or organ-
ic solvents containing xylene or acetone.
The external painted shell can be kept clean with a humid cloth.
Prevent liquids from entering the unit while cleaning or during operation.
Please refer to your local supplier for regular service inspection which should be planned every six
(6) months aver installation.
Removal of marking dust
.
Procedure:
•
Always make sure the unit is powered OFF
•
Remove the hoppers carousel and the convex lid (4 screws)
•
With a brush remove any dust residual that might have deposited on the chute.
•
Replace the lid and the carousel.
•
Power ON the unit and press the Emergency button that will turn RED.
•
Power on the PC.
•
Double click the VEGA Icon and start marking.
LASER LENS
Cleaning every 6 months.
Procedure:
•
First remove the right-side shell by releasing two screws. Remove the instrument shell.
•
Prepare a solution of 5 to 8% acetic acid.
•
Use a 4 X 4 gauze pad, moisten the gauze in the solution.
•
Gently Clean the laser LENS
•
Dry the lens and the metallic parts surrounding the LENS.
Replace the shell back onto the instrument, and test the marking quality by printing several cas-
settes
Mechanical test
every 6months.
Procedure:
•
Follow the same procedures as explained for LASER LENS Cleaning to remove the in-
strument shell.
•
Refer to our “Service manual “for details.