OPERATING INSTRUCTIONS
3-14
Handle the Hall probe with care. Do not bend the stem or
apply pressure to the probe tip as damage may result.
In other situations the user may want the probe to be exposed to a
specific magnetic field during the zeroing process so that all future
readings do not include that reading (such as the earth’s field).
This is possible with the following restrictions:
1) The external field must not exceed 30 mT (300 G).
2) The field must be stable during the zeroing process. It should
not contain alternating (ac) components.
AUTOMATIC ZERO FUNCTION
The meter provides two methods to zero the probe. The first is
completely automatic. Prepare the probe for zeroing, then rotate
the function selector to the ZERO position. The “ZERO” legend
will flash and actual dc flux density readings will appear on the
display. The meter will select the lowest range regardless of which
range was in use prior to using the ZERO function. Recall that the
maximum flux density level that can be zeroed is 30 mT (300 G).